Thermo Scientific Multifuge X1 Centrifuge
Minimum / Maximum RPM 300 / 15,200
Maximum Capacity 4 x 400mL
- Intuitive control panel with a digital display for detailed operation monitoring.
- Adjust speed, time, and temperature during the run using a convenient keyboard.
- One-finger lid closing and locking, with optional automatic lid opening.
- Easy push-button rotor installation and exchange saves time between applications.
- Easy-to-clean chamber for a healthier workspace and extended unit lifespan.
- Glove-friendly, one-handed operation for open/close functionality.
- Advanced rotor management for optimized acceleration, braking, and load balancing.
- Compact, ergonomic design for efficient use of space and comfortable handling.
- Energy-efficient design saves up to 40% energy on standard protocols.
The Thermo Scientific Multifuge X1 Centrifuge
features a maximum speed of 15,200 rotations per minute and a maximum relative
centrifugal force of 25,830 times gravity. An onboard LCD display makes reading
information easy and entering parameters a breeze. The X1 has a maximum timer
run time of 99 hours and 59 minutes and also offers the ability to run indefinitely.
With a maximum capacity of 4 x 400 milliliters, the Thermo Scientific Heraeus Multifuge X1 Centrifuge offers
an extremely high throughput for a relatively compact centrifuge. The Auto-Lock III Rotor System features a s
ecure locking system allowing for easy push-button installation and exchange of rotor which saves time in switch between applications. We may have different adapters available depending on the rotor chosen with this centrifuge. If the selected rotor includes adapters that will not work for your application, please inquire.
Please confirm the selected rotor meets your application needs and request a different rotor, if necessary. Available rotors can be found in our Rotor Compatibility Guide.
The Thermo Scientific Multifuge X1 Centrifuge includes:
- > Selected Rotor Above
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Multifuge X1 Centrifuge Specifications
Maximum Speed
|
15,200 RPM
|
Min Speed
|
300 rpm
|
Max Capacity
|
4 x 400mL
|
Maximum RCF
|
25,830 x G
|
Timer
|
99 hours, 59 minutes, plus HOLD
|
Program Storage
|
99 Programs
|
Dimensions
|
14.2 x 17.3 x 23.8 in. (36 x 44 x 60.5cm)
|
Weight without Rotor
|
57.5 kg
|
Controller Type
|
Microprocessor
|
Refrigeration
|
No Refrigeration
|
Click Here to Expand Rotor Specifications
TX-200 Rotor Specs
|
Capacity
|
4 x 180 mL
|
Max RPM
|
5,500
|
Max RCF
|
5,580 x g
|
Fiberlite F-15-8x50Cy Rotor Specs
|
Capacity
|
8 x 50 mL Conical tubes
|
Max RPM
|
14,500
|
Max RCF
|
24,652 x g
|
Fiberlite F-15-6x100Y Rotor Specs
|
Capacity
|
6 x 100 mL
|
Max RPM
|
15,000
|
Max RCF
|
24,652 x g
|
M-20 Microplate Rotor Specs
|
Capacity
|
2 x 3 Standard or 2 x 1 Deepwell Plates
|
Max RPM
|
4,000
|
Max RCF
|
2,272 x g
|
TX-400 Rotor Specs
|
Capacity
|
4 x 400 mL
|
Max RPM
|
5,000
|
Max RCF
|
4,696 x g
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task |
Completed? |
Remarks? |
Confirm power supply output voltages match specified requirements |
Completed |
|
Measure and verify voltages at PCB test points for accuracy |
Completed |
|
Confirm rotor compatibility and detection for proper identification |
Completed |
|
Inspect exterior for cosmetic or structural damage |
Completed |
|
Check door mechanism alignment for secure opening and closing |
Completed |
|
Inspect and replace gas strut with OEM part as needed |
Completed |
|
Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
Replace and confirm functionality of all lid lock switches |
Completed |
|
Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
Test compressor (if applicable) for proper operation and cooling |
Completed |
|
Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
Check rotor motor for stability, noise, and integrity |
Completed |
|
Inspect and clear condensation drain to prevent buildup |
Completed |
|
Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|