Thermo Scientific ST40R Refrigerated Centrifuge
- Speeds Up To 15,200 RPM
- Includes TX-750 Rotor, Buckets & Adapters
- Fully Recertified
- 90-Day Full Parts & Labor Warranty
Go to Product Details
Thermo Scientific Megafuge 40R Refrigerated Centrifuge
Maximum Capacity 4 x 1000 mL
Temperature Range -10°C to 40°C
- Benchtop centrifuge designed with a user-friendly control panel for easy pre-setting of parameters
- Capable of reaching maximum speeds of 15,200 RPM and a maximum RCF of 25,314 x g
- Features a broad temperature range from -10 to +40°C, suitable for various sample types
- Refrigerated to maintain stable temperatures during high-speed operations
- Equipped with a maintenance-free induction motor for quiet, low-vibration, and durable performance
- Large capacity of 4 x 1000mL, accommodating substantial sample volumes
- Includes digital display and microprocessor controller for precise and convenient operation
- Provides program storage for 6 programs with direct access, and a runtime up to 9 hours 99 minutes plus HOLD
The Thermo Scientific Megafuge 40R Refrigerated Centrifuge is a benchtop centrifuge designed with the user in mind.
The user-friendly control panel makes it easy to pre-set the speed, RCF value, running time, temperature, and running profile. The set speed is reached within seconds. The Megafuge 40 can reach maximum speeds of 15,200 RPM and maximum RCF of 25,314 x g and has a t emperature range from -10 to +40C. The maintenance-free induction motor ensures quiet and low-vibration operation even at high speeds, and guarantees a very long lifetime. We may have different adapters available depending on the rotor chosen with this centrifuge. If the selected rotor includes adapters that will not work for your application, please inquire.
All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
The Thermo Scientific Megafuge 40R Refrigerated Centrifuge includes:
- > Rotor Selected Above
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Megafuge 40R Refrigerated Centrifuge Specifications
Capacity
|
4 x 1000mL
|
Max. RCF
|
25,314 x g
|
Max. Speed
|
15,200rpm (fixed angle rotor)
|
Refrigerated
|
Yes
|
Display
|
Digital
|
Height w/ Open Cover
|
35.4 in. (90 cm)
|
Wattage
|
1400w
|
Electrical Requirements
|
120V 60Hz
|
Controller Type
|
Microprocessor
|
Dimensions (H x W x D)
|
14.2 x 29.3 x 26.4 in. (36 x 74.5 x 67cm)
|
Drive System
|
Direct, brushless induction low-profile motor
|
Net Weight
|
255.7 lb. (116kg)
|
Program Storage
|
6 programs, direct access
|
Run Time
|
9 hr. 99 min. plus HOLD
|
Temperature Range
|
-10 to +40C
|
Click Here to Expand Rotor Specifications
TX-750 Rotor Specs
|
Capacity
|
4 x 750 mL
|
Max RPM
|
4,300
|
Max RCF
|
4,031 x g
|
Bioliner Rotor Specs
|
Capacity
|
4 x 750 mL
|
Max RPM
|
3,500
|
Max RCF
|
2,739 x g
|
Fiberlite F-15-6x100Y Rotor Specs
|
Capacity
|
6 x 100 mL
|
Max RPM
|
15,000
|
Max RCF
|
24,652 x g
|
M-20 Microplate Rotor Specs
|
Capacity
|
2 x 3 Standard or 2 x 1 Deepwell Plates
|
Max RPM
|
4,00
|
Max RCF
|
2,272 x g
|
TX-1000 Rotor Specs
|
Capacity
|
4 x 1000 mL
|
Max RPM
|
4,200
|
Max RCF
|
4,122 x g
|
Thermo Scientific ST40R Refrigerated Centrifuge
- Speeds Up To 15,200 RPM
- Includes TX-750 Rotor, Buckets & Adapters
- Fully Recertified
- 90-Day Full Parts & Labor Warranty
Go to Product Details
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task |
Completed? |
Remarks? |
Confirm power supply output voltages match specified requirements |
Completed |
|
Measure and verify voltages at PCB test points for accuracy |
Completed |
|
Confirm rotor compatibility and detection for proper identification |
Completed |
|
Inspect exterior for cosmetic or structural damage |
Completed |
|
Check door mechanism alignment for secure opening and closing |
Completed |
|
Inspect and replace gas strut with OEM part as needed |
Completed |
|
Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
Replace and confirm functionality of all lid lock switches |
Completed |
|
Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
Test compressor (if applicable) for proper operation and cooling |
Completed |
|
Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
Check rotor motor for stability, noise, and integrity |
Completed |
|
Inspect and clear condensation drain to prevent buildup |
Completed |
|
Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|