Thermo Scientific MaxQ 6000 Refrigerated Incubator Shaker
The Thermo Scientific MaxQ 6000 Orbital Incubator Shaker Model: SHKE6000-7 (4353) combines orbital shaking technology with incubation and easy-to-use digital controls for a versatile shaker ideal for cell culture, solubility studies, extraction procedures, protein expression, and many other applications. The MaxQ 6000 features shaking speeds from 15 to 500 RPM around a 0.75 inch orbit with a temperature range of 10C above ambient to 80C. The Thermo Scientific MaxQ 6000 Orbital Incubator Shaker is powered by a triple-eccentric drive mechanism with three shafts supporting the platform for smooth and uniform operation, excellent product yields, routine handling of heavy workloads, and continuous 24-hour operation. A spacious 18 x 18 inch universal platform is also included, which allows for higher vessel capacity for up to six 2L flasks. The MaxQ 6000's clear, scratch-resistant window allows for unobstructed sample viewing without disturbing the chamber. The PID temperature controller provides 0.1C temperature accuracy at 37C, and an independent over temperature feature provides a backup temperature controller for added safety. This incubator shaker will be fully tested to perform at factory specifications before being shipped.
The Thermo Scientific MaxQ 6000 Orbital Incubator Shaker includes:
- > 18 x 18 inch Universal Platform
- > Shelf
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific MaxQ 6000 Orbital Incubator Shaker Specifications
Speed Range
|
15 to 500 RPM (unstacked)
15 to 300 RPM (stacked)
|
Speed Accuracy
|
1 RPM
|
Temperature Range
|
15C below Ambient to 80C
|
Temperature Uniformity
|
0.5 at 37C
|
Temperature Accuracy
|
0.1C at 37C
|
Timer
|
Continuous or timed operation from 0.1 to 999 minutes/hours
|
Platform Size
|
18 x 18 in
|
Orbit
|
0.75 in (1.9 cm)
|
Chamber Dimensions (L x W x H)
|
22 x 21 x 27 in (55.9 x 53.3 x 68.6 cm)
|
Overall Dimensions (L x W x H)
|
33 x 27.5 x 40.5 in (83.8 x 69.8 x 102.8 cm)
|
Weight
|
330 lbs. (150 kg)
|
Electrical Requirements
|
120V, 60Hz, 11A, 1100W
|
Thermo Scientific MaxQ 6000 ManualClick Here to View the Technical Specs for the Thermo Scientific MaxQ 6000
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Examine chamber glass for cracks or chips, replace as needed | Completed | |
Inspect door seals for wear, replace to ensure tight seal | Completed | |
Check display brightness, replace dim or dead pixels | Completed | |
Test all buttons for full functionality and responsiveness | Completed | |
Replace platform assembly bearings for reliable motion | Completed | |
Inspect drive belt for wear, replace if torn or cracked | Completed | |
Examine wiring for frays or breakages, repair as needed | Completed | |
Verify PCB test points, replace faulty components | Completed | |
Check drive belt tension for proper alignment | Completed | |
Confirm compressor functionality, clean filter and fins | Completed | |
Calibrate internal temperature with NIST probe | Completed | |
Calibrate platform RPM with NIST-traceable tachometer | Completed | |
Conduct cosmetic inspection for appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|