Thermo Scientific MaxQ 4000 Orbital Incubator Shaker
Temperature Range Ambient +10°C to 60°C
Temperature Uniformity
0.5 at 37C
- Combines orbital shaking with benchtop design for versatility in various applications like cell culture and protein expression
- Shaking speeds range from 40 to 400 RPM with a 0.75-inch orbit, adaptable for multiple research needs
- Temperature range from 10°C above ambient to 60°C, suitable for temperature-sensitive applications
- Spacious 18 x 18 inch platform holds up to six 2L flasks, accommodating larger vessel capacities
- Clear, scratch-resistant lid provides unobstructed sample viewing without disrupting chamber conditions
- PID temperature controller ensures precise 0.1°C accuracy at 37°C for consistent environmental conditions
- Backup over-temperature feature offers additional safety by independently controlling temperature
- Temperature uniformity of 0.5°C at 37°C, maintaining stable conditions across the platform
- Timer options for continuous or timed operation from 1 to 60 minutes, ideal for varied experimental durations
- Compact design with dimensions of 32 x 22.5 x 22 inches, fitting comfortably on most benchtops
The Thermo Scientific MaxQ 4000 combines orbital shaking technology with benchtop design for a versatile shaker ideal for cell culture, solubility studies, extraction procedures, protein expression, and many other applications. The MaxQ 4000 features shaking speeds from 40 to 400 RPM around a 0.75 inch orbit with a temperature range of 10C above ambient to 60C. Spacious 18 x 18 inch platform allows for higher vessel capacity for up to six 2L flasks. The MaxQ 4000's clear, scratch-resistant lid allows for unobstructed sample viewing without disturbing the chamber. PID temperature controller provides 0.1C temperature accuracy at 37C, and an independent over temperature feature provides a backup temperature controller for added safety. This incubator shaker will be fully tested to perform at factory specifications before being shipped.
The Thermo Scientific MaxQ 4000 Orbital Incubator Shaker includes:
- > 18 x 18 in Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific MaxQ 4000 Orbital Incubator Shaker Specifications
Speed Range
|
40 to 400 RPM
|
Speed Accuracy
|
10 RPM
|
Temperature Range
|
Ambient +10 to 60C
|
Temperature Uniformity
|
0.5 at 37C
|
Temperature Accuracy
|
0.1 at 37C
|
Timer
|
Continuous or timed operation from 1 to 60 minutes
|
Dimensions (L x W x H)
|
32 x 22.5 x 22 in (81.3 x 57.2 x 55.9 cm)
|
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Examine chamber glass for cracks or chips, replace as needed | Completed | |
Inspect door seals for wear, replace to ensure tight seal | Completed | |
Check display brightness, replace dim or dead pixels | Completed | |
Test all buttons for full functionality and responsiveness | Completed | |
Replace platform assembly bearings for reliable motion | Completed | |
Inspect drive belt for wear, replace if torn or cracked | Completed | |
Examine wiring for frays or breakages, repair as needed | Completed | |
Verify PCB test points, replace faulty components | Completed | |
Check drive belt tension for proper alignment | Completed | |
Confirm compressor functionality, clean filter and fins | Completed | |
Calibrate internal temperature with NIST probe | Completed | |
Calibrate platform RPM with NIST-traceable tachometer | Completed | |
Conduct cosmetic inspection for appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|