Thermo Scientific Heratherm OMH60 Advanced Protocol Oven, Cat. #: 51028124
The Thermo Scientific Heratherm OMH60 Advanced Protocol Oven, Cat. #: 51028124 is a laboratory oven that is designed for efficiency and can be used with multiple different applications. The oven features a timer that allows for preset on/off times for each day or based on real-time hours and minutes. This functionality aims to optimize energy usage. The oven can operate at temperatures up to 330°C and is designed to maintain consistent temperature uniformity. One of its distinguishing features is an adjustable fan speed that can be set to five different settings, tailored to specific applications, ranging from rapid drying to maintaining temperature stability.
The programmable controller can manage temperature ramps and dwells and can store up to 10 different programs, each with 10 steps. It's also equipped with an electronically controlled fan speed and damper position. The oven includes an access port of 0.71 inches in diameter for introducing independent sensors for data monitoring. Furthermore, it offers a calibration routine to ensure ongoing temperature accuracy and a boost function for quicker heating, reducing the need for continuous operation.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The Thermo Scientific Heratherm OMH60 Advanced Protocol Oven, Cat. #: 51028124 includes:
- > 2 Shelves
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Heratherm OMH60 Advanced Protocol Oven, Cat. #: 51028124 Specifications
Temperature Range
|
Ambient +10°C to 330°C
|
Temperature Uniformity
|
± 1.5°C at 150°C
|
Temperature Stability
|
± 0.2°C at 150°C
|
Volume
|
2.2 cu. ft. (61L)
|
Heat Up Time
|
18 minutes to 150°C
|
Number of Shelves
|
2
|
Maximum Shelf Load
|
55 lb. (25 kg)
|
Internal Dimensions (W x H x D)
|
13.9 x 20 x 13.5 in. (35.4 x 50.8 x 34.3 cm)
|
External Dimensions (W x H x D)
|
20.9 x 29.7 x 22.2 in. (53 x 75.5 x 56.9 cm)
|
Weight
|
113 lb. (51.3 kg.)
|
Electrical Requirements
|
120V, 60Hz, 12A, 1440W
|
- Verify power supply and PCB outputs are within specified voltage ranges
- Inspect display for any dim or dead pixels, replacing as needed to ensure visibility
- Confirm full functionality and responsiveness of all control buttons
- Examine all internal wiring for frays, cuts, or damage, ensuring secure and reliable connections
- Inspect door seals and gaskets for integrity, replacing any compromised components
- Evaluate door handle and locking mechanism for wear or issues, replacing if performance is impaired
- Inspect and clean internal fan; replace if performance does not meet requirements
- Calibrate set temperature using a NIST-traceable probe for precise accuracy
- Confirm functionality of over-temperature alarm and indicator light
- Verify internal outlet functionality, if applicable, to ensure it meets operational standards
- Inspect internal lighting for full functionality and brightness, if applicable
- Perform a cosmetic inspection to confirm the oven’s exterior and interior meet quality standards
Technician Inspection and Recertification Checklist: Ovens
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Verify power supply and PCB outputs are within specified voltage ranges
|
Completed
|
|
Inspect display for dim or dead pixels; replace if needed
|
Completed
|
|
Confirm full functionality of all control buttons
|
Completed
|
|
Examine internal wiring for damage; ensure secure connections
|
Completed
|
|
Inspect door seals and gaskets, replacing any compromised components
|
Completed
|
|
Evaluate door handle and locking mechanism, replacing if worn
|
Completed
|
|
Inspect and clean internal fan; replace if performance is insufficient
|
Completed
|
|
Calibrate set temperature using a NIST-traceable probe
|
Completed
|
|
Confirm over-temperature alarm and indicator light functionality
|
Completed
|
|
Verify internal outlet functionality, if applicable
|
Completed
|
|
Inspect internal lighting for full functionality and brightness
|
Completed
|
|
Perform cosmetic inspection to confirm quality standards of exterior and interior
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|