Popular Alternative Choice
VWR Gravity Convection Incubator, Cat. #: 89511-422
- VWR Incubator offers Ambient +5°C to 75°C, with ± 0.6°C uniformity at 37°C.
- Features: PID control, RS-232 interface, and overtemperature alarm.
- Includes 2 shelves, calibration report, and 180-day warranty.
- Dimensions: 18.3x27.9x23.2 in. (internal) and 25.2x36.2x29.1 in. (external).
Go to Product Details
Fully Recertified!
Thermo Scientific Heratherm IGS180 Incubator
The Thermo Scientific Heratherm IGS180 Incubator utilizes gravity convection for gentle air flow and minimal sample drying. This incubator provides a temperature range from ambient +5C to 75C with a uniformity of 0.6C and stability of 0.2C. With an intuitive interface and large vacuum fluorescent read-out, the Heratherm incubator is incredibly easy-to-use and user friendly. Corrosion-resistant stainless steel interior and internal glass door allow for long life and ease of viewing samples.
The Thermo Heratherm incubator is designed for complete sample safety, including applications in pharmaceutical, medical, food, and research laboratories.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The Thermo Scientific Heratherm IGS180 Incubator includes:
- > 2 Shelves
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Heratherm IGS180 Incubator Specifications
Temperature Range
|
Ambient +5C to 75C
|
Temperature Uniformity
|
0.6C
|
Temperature Stability
|
0.2C
|
Volume
|
194 L (6.9 cu ft)
|
Interior Dimensions (D x W x H)
|
58.9 x 46.4 x 70.8 cm (23.2 x 18.3 x 27.9 in)
|
Exterior Dimensions (L x W x H)
|
73.8 x 64 x 92 cm (29.1 x 25.2 x 36.2 in)
|
Electrical Requirements
|
120V, 60Hz, 6A, 720W
|
Thermo Heratherm Incubator Operator's Manual
Click Here to View the Technical Specs for the Heratherm IGS180
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|