Thermo Scientific Heracell VIOS 250i CO2 Incubator
The Thermo Scientific Heracell VIOS 250i CO2 Incubator is specifically designed to deliver the performance reliability, ease of operation, and value required to support a range of culturing needs from basic research to the demanding, leading-edge applications. This incubator is reliable and easy to use while featuring fast recovery characteristics and a convenient touchscreen user interface for easily controlling the temperature from Ambient +3°C to 55°C. The Thermo Scientific iCAN™ touchscreen simplifies operation and enables rapid access of important information for each critical parameter in the incubator. This can provide trend analysis for convenient evaluation of the unit’s performance. The touchscreen allows for alert notifications such as when gas supply is low, or when there are changes to the cell culture environment. The Heracell VIOS 250i incorporates THRIVE active airflow technology, providing faster recovery and uniformity for consistent results. An in-chamber fan gently and evenly distributes clean, humidified air throughout the chamber ensuring all cells experience the same conditions without the threat of desiccation. Cells experience total recovery of all critical growth parameters in less than 10 minutes following a 30 second door opening. Also featured is in-chamber HEPA filtration, which continuously filters the entire chamber air volume every 60 seconds. ISO 5 clean room air quality is achieved in the culture chamber within 5 minutes. Continuous protection is provided by minimizing the ability for contaminants to settle on interior surfaces. The exclusive "Steri-Run" high temperature sterilization cycle included with this incubator reaches 180°C on all chamber surfaces and is independently proven to achieve total sterilization and a 12 log Sterility Assurance Level (SAL). With the push of a button, the simple overnight routine provides fast, easy elimination of microbial contaminants and eliminates the need for separate autoclaving of parts. Additionally, the 250i’s protected humidity reservoir boasts a directly heated water reservoir which provides stable high relative humidity levels, preventing culture desiccation. This allows for up to five times faster humidity recovery than traditional water pan designs. All of our equipment is fully tested to perform at factory specifications before being shipped.
The Thermo Scientific Heracell VIOS 250i CO2 Incubator includes:
- > 2 x Shelves
- > Integrated Humidity Reservoir
- > Stacking Kit
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Heracell VIOS 250i CO2 Incubator Specifications
Temperature Range
|
Ambient +3°C to 55°C
|
Temperature Uniformity
|
± 0.3°C at 37°C
|
Temperature Stability
|
± 0.1°C at 37°C
|
CO2 Range
|
1 to 20%
|
CO2 Control Deviation
|
± 0.1%
|
CO2 Accuracy
|
± 0.3%
|
CO2 Sensor Type
|
TC
|
Required Gas Pressure
|
Minimum of 0.8 bar to maximum of 1 bar (12 to 15 PSI)
|
Relative Humidity
|
> 93% at 37°C
|
Volume
|
9.0 cu ft. (255L)
|
Sterilization Cycle Temperature
|
180°C on all internal surfaces
|
Sterilization Cycle Duration
|
Under 12 hours
|
Access Port Diameter
|
1.6 in. (42 mm)
|
Data Outputs
|
Remote alarm contacts and USB
|
Shelf Dimensions (W x D)
|
16.7 x 18.3 in. (42.3 x 46.5 cm)
|
Internal Dimensions (W x H x D)
|
23.9 x 26.4 x 24.8 in. (60.7 x 67 x 62.9 cm)
|
External Dimensions (W x H x D)
|
30.7 x 38.2 x 37.2 in. (78 x 97 x 94.5 cm)
|
Weight
|
215 lb. (97.5 kg.)
|
Electrical Requirements
|
120V, 50/60 Hz
|
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|