Thermo Scientific Heating Digital Shaking Drybath with 1.5/2.0ml Block, Cat. #: 88880027
The Thermo Scientific Heating Digital Shaking Drybath, Cat. #: 88880027, offers a combination of drybath and shaker functionalities. With a speed range of 150 to 1,500 RPM (150 to 1,200 RPM with included block), it can handle different block types, affecting the maximum RPM achievable. Its temperature capabilities range from Ambient +5°C to 100°C, maintaining a uniformity of ± 0.5°C and stability of ± 0.15°C at 80°C. The unit features a PID control system for consistent shaking and accurate temperature regulation. It can store over 10 programs with up to 10 steps each, and offers an interval shaking function, enabling both shaking and stationary modes in the same program. Timing can be initiated immediately or once the desired temperature is reached. Designed for flexibility, it supports easily interchangeable blocks of various sizes.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Thermo Scientific Heating Digital Shaking Drybath with 1.5/2.0ml Block, Cat. #: 88880027 includes:
- > 24 x 1.5/2.0ml Block
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Heating Digital Shaking Drybath with 1.5/2.0ml Block, Cat. #: 88880027 Specifications
Speed Range
|
150 to 1,500 RPM (150 to 1,200 RPM with included block)
|
Temperature Range
|
Ambient +5°C to 100°C
|
Temperature Uniformity
|
± 0.5°C at 80°C
|
Temperature Stability
|
± 0.15°C at 80°C
|
Temperature Increments
|
± 0.1°C
|
Heating/Cooling Rate
|
Heating: 5°C/minute
|
Orbit
|
2 mm
|
Timer
|
1 minute to 99 hours and 59 minutes, or continous
|
Display
|
Digital
|
Maximum Heater Output
|
360W
|
Dimensions (W x D x H)
|
10.9 x 13.1 x 6.7 in. (27.6 x 33.4 x 17 cm)
|
Weight
|
18.3 lb. (8.3 kg)
|
Electrical Requirements
|
120V, 60Hz, 3.4A
|
- Confirm stability and accuracy of power supply and PCB voltage outputs
- Inspect display integrity, ensuring pixel clarity and replacing any degraded or non-functional elements
- Verify complete functionality and responsiveness of all control buttons, switches, and knobs
- Conduct thorough inspection of internal wiring and all electrical connections for continuity; replace or secure any degraded or substandard components
- Examine block holder platform for alignment, obstructions, or error-inducing irregularities
- Perform detailed inspection of motor bearings, assessing for wear or misalignment, and replace if performance deviates from specifications
- Calibrate rotational speed precisely using a NIST-traceable tachometer for compliance with operational standards
- Calibrate temperature accuracy using a NIST-traceable temperature probe, validating readings across the operational range
- Ensure timer precision by validating functionality and consistency against set parameters
- Conduct a comprehensive cosmetic inspection, verifying all surfaces and external components meet quality standards
Technician Inspection and Recertification Checklist: Thermo Mixer
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Confirm stability and accuracy of power supply and PCB voltage outputs
|
Completed
|
|
Inspect display integrity, ensuring pixel clarity and replacing any degraded elements
|
Completed
|
|
Verify functionality and responsiveness of all control buttons, switches, and knobs
|
Completed
|
|
Inspect internal wiring and connections for continuity, replacing or securing degraded components
|
Completed
|
|
Examine block holder platform for alignment, obstructions, or irregularities
|
Completed
|
|
Inspect motor bearings for wear or misalignment, replacing if necessary
|
Completed
|
|
Calibrate rotational speed using an NIST-traceable tachometer
|
Completed
|
|
Calibrate temperature with an NIST-traceable probe, validating accuracy
|
Completed
|
|
Ensure timer precision by validating against set parameters
|
Completed
|
|
Conduct cosmetic inspection for quality standard compliance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|