Thermo Scientific Digital Microplate Shaker, Cat. #: 88882005
The Thermo Scientific Digital Microplate Shaker, Cat. #: 88882005 is designed to provide stable and accurate shaking of microplates, making it ideal for various applications such as immunoassay, screening, magnetic bead, and other specialized processes. This shaker is capable of handling up to four 96-well microplates simultaneously, thanks to its standard and enlarged platform. It boasts a wide speed range from 150 to 1,200 RPM, ensuring speed accuracy of ±1% up to 299 RPM and ±2% between 300 to 1,200 RPM.
Equipped with a timer that can be set from 1 minute to 99 hours and 59 minutes or operated continuously, the Thermo Scientific Digital Microplate Shaker caters to a variety of experimental needs. The shaker features a 2.5 mm orbit, a clear LED display, and can accommodate a maximum load of 8.8 lb. (4 kg). The platform measures 10.8 x 9.3 inches (27.5 x 23.5 cm), allowing for efficient and precise shaking for a variety of laboratory needs.
The Thermo Scientific Digital Microplate Shaker, Cat. #: 88882005 includes:
- > 10.8" x 9.3" Microplate Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator’s Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Digital Microplate Shaker, Cat. #: 88882005 Specifications
Speed Range
|
150 to 1,200 RPM
|
Speed Accuracy
|
± 1% up to 299 RPM
± 2% 300 to 1,200 RPM
|
Timer
|
1 minutes to 99 hours 59 minutes or continuous
|
Orbit
|
2.5 mm (0.098 in.)
|
Display
|
LED
|
Maximum Load
|
8.8 lb. (4 kg)
|
Microplate Capacity
|
4 x 96-well microplates
|
Platform Size (L x W)
|
10.8 x 9.3 in. (27.5 x 23.5 cm)
|
Dimensions (D x W x H)
|
15.1 x 10.2 x 5.7 in. (38.3 x 26 x 14.5 cm)
|
Weight
|
28.2 lb. (12.8 kg)
|
Electrical Requirements
|
100-240V, 50/60Hz
|
- Validate that the power supply and PCB deliver stable, correct voltage outputs across all necessary connections, ensuring reliable operation
- Thoroughly inspect the display for any dim, flickering, or dead pixels; replace affected elements to maintain full readability and operational clarity
- Test the full range of button functionality, confirming responsive, accurate input recognition across all settings
- Conduct a detailed inspection of bearings and bearing housing, identifying any worn or damaged bearings, and replace as needed to prevent operational disruption
- Examine the drive belt for signs of cracking, fraying, or material breakdown; replace as required to ensure sustained performance
- Measure and adjust motor drive belt tension to optimal specifications, securing reliable and efficient power transmission
- Verify accuracy of the onboard timer, confirming proper functionality and replacing if it deviates from expected performance
- Calibrate shaker rotation speed with a NIST-traceable tachometer, validating precision within specified operational limits
- Run the shaker continuously for an 8-12 hour testing period, closely observing the shaking platform for smooth, consistent motion, while monitoring for any error indicators such as knocking, skipping, or unusual vibrations
- Perform a comprehensive cosmetic inspection to confirm the device’s exterior meets aesthetic and quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Shakers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed
|
Completed
|
|
Test button functionality for responsiveness and accuracy across all settings
|
Completed
|
|
Inspect bearings and housing for wear; replace as needed to prevent disruption
|
Completed
|
|
Examine drive belt for cracking, fraying, or material breakdown; replace as required
|
Completed
|
|
Measure and adjust motor drive belt tension to optimal specifications
|
Completed
|
|
Verify onboard timer accuracy; replace if it deviates from expected performance
|
Completed
|
|
Calibrate shaker rotation speed using a NIST-traceable tachometer
|
Completed
|
|
Run 8-12 hour continuous test; observe platform for smooth, consistent motion
|
Completed
|
|
Perform comprehensive cosmetic inspection for quality standards compliance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|