|
|
|
|
Thermo Scientific Compact Digital Rocker, Cat. #: 88880019
The Thermo Scientific Compact Digital Rocker, Cat. #: 88880019 is a highly efficient and reliable tool designed for gentle sample agitation in a range of laboratory applications, including ELISA, protein synthesis, staining/de-staining of gels, and antibody staining. This rocker offers a speed range of 5 to 100 RPM with an accuracy of ±5% RPM, and its PID control ensures a consistent and smooth rocking motion. The tilt angle can be easily adjusted between 7° to 13°, catering to various experimental requirements.
Featuring a compact design, this rocker is ideal for use in incubators (non-CO2), chambers, and refrigerators up to 60°C. The platform measures 10.87 x 10.87 inches (27.6 x 27.6 cm) and can handle a maximum load of 11.02 lb. (5 kg), making it suitable for accommodating up to six well-plates. It comes equipped with a Vacuum Fluorescent Display (VFD) for clear visibility of settings. Additionally, the timer can be set for continuous operation or from 1 minute to 99 hours and 59 minutes, providing flexibility for time-dependent and low-dependence sensitive applications.
The Thermo Scientific Compact Digital Rocker, Cat. #: 88880019 includes:
- > 10.87" x 10.87" Dimpled Rubber Mat Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator’s Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Compact Digital Rocker, Cat. #: 88880019 Specifications
Speed Range
|
5 to 100 RPM
|
Speed Accuracy
|
± 5% RPM
|
Tilt Angle Range
|
7° to 13°
|
Timer
|
1 min. to 99 hrs. 59 min. or continuous
|
Display
|
Vacuum Fluorescent Display (VFD)
|
Maximum Load
|
11.02 lb. (5 kg)
|
Platform Size (W x D)
|
10.87 x 10.87 in. (27.6 x 27.6 cm)
|
Dimensions (W x D x H)
|
10.87 x 12.6 x 5.91 in. (27.6 x 32 x 15 cm)
|
Weight
|
8.16 lb. (3.7 kg)
|
Electrical Requirements
|
100-240V, 50/60Hz
|
- Validate that the power supply and PCB deliver stable, correct voltage outputs across all necessary connections, ensuring reliable operation
- Thoroughly inspect the display for any dim, flickering, or dead pixels; replace affected elements to maintain full readability and operational clarity
- Test the full range of button functionality, confirming responsive, accurate input recognition across all settings
- Conduct a detailed inspection of bearings and bearing housing, identifying any worn or damaged bearings, and replace as needed to prevent operational disruption
- Examine the drive belt for signs of cracking, fraying, or material breakdown; replace as required to ensure sustained performance
- Measure and adjust motor drive belt tension to optimal specifications, securing reliable and efficient power transmission
- Verify accuracy of the onboard timer, confirming proper functionality and replacing if it deviates from expected performance
- Calibrate shaker rotation speed with a NIST-traceable tachometer, validating precision within specified operational limits
- Run the shaker continuously for an 8-12 hour testing period, closely observing the shaking platform for smooth, consistent motion, while monitoring for any error indicators such as knocking, skipping, or unusual vibrations
- Perform a comprehensive cosmetic inspection to confirm the device’s exterior meets aesthetic and quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Shakers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed
|
Completed
|
|
Test button functionality for responsiveness and accuracy across all settings
|
Completed
|
|
Inspect bearings and housing for wear; replace as needed to prevent disruption
|
Completed
|
|
Examine drive belt for cracking, fraying, or material breakdown; replace as required
|
Completed
|
|
Measure and adjust motor drive belt tension to optimal specifications
|
Completed
|
|
Verify onboard timer accuracy; replace if it deviates from expected performance
|
Completed
|
|
Calibrate shaker rotation speed using a NIST-traceable tachometer
|
Completed
|
|
Run 8-12 hour continuous test; observe platform for smooth, consistent motion
|
Completed
|
|
Perform comprehensive cosmetic inspection for quality standards compliance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|