|
|
|
|
Thermo Scientific 4140 CO2 Water Jacketed Incubator w/ O2 Control
The Thermo Scientific 4140 CO2 Water Jacketed Incubator w/ O2 Control includes reliable water jacket technology with an effortless Thermo Scientific iCAN touchscreen. Dependable temperature stability and HEPA clean room air purity are paired with the simplest ways to control and monitor the incubator, ensuring protection for important cell cultures. The 4140 allows users to control temperature from Ambient +5°C to 55°C, control CO2 from 0 to 20%, and control O2 from 1 to 20%, providing a combination that’s truly hard to beat. The 6.5 cu. ft. (184.1L) high-capacity chamber provides outstanding culturing space and features a unique triple wall construction, providing an outstanding temperature uniformity ± 0.2°C at 37°C. The Thermo Scientific 4140 features in-chamber HEPA air filtration, which minimizes the risk of airborne contaminants entering the incubator from multiple door openings. The HEPA filtration reaches ISO 5 clean room air quality in 5 minutes by filtering the chamber air every 60 seconds. The bright, door-mounted iCAN touchscreen interface provides complete data visibility to monitor the incubator. This interface specifically features on-board error and data logging for monitoring performance, interactions and alarm conditions. The 4140 also provides enhanced flexibility with complete O2 control as many cell cultures thrive best in CO2 incubators with controlled levels of oxygen. Users can select an O2 option from 1 to 20% to simulate physiological or hypoxic environments. All of our equipment is fully tested to perform at factory specifications before being shipped.
The Thermo Scientific 4140 CO2 Water Jacketed Incubator w/ O2 Control includes:
- > 3 Shelves
- > Drain Tray
- > New HEPA Filter
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific 4140 CO2 Water Jacketed Incubator w/ O2 Control Specifications
Temperature Range
|
Ambient +5°C to 55°C
|
Temperature Uniformity
|
± 0.2°C at 37°C
|
Temperature Control
|
± 0.1°C
|
Volume
|
6.5 cu. ft. (184.1L)
|
CO2 Range
|
0 to 20%
|
CO2 Control
|
± 0.1%
|
CO2 Sensor Type
|
IR (Infrared)
|
O2 Range
|
1 to 20%
|
O2 Control
|
± 0.1%
|
O2 Overshoot
|
± 0.2%
|
Inlet Pressure
|
15 PSIG, (1.034 bars), ± 5 PSIG
|
Relative Humidity
|
Ambient to 95% at 37°C
|
Water Jacket Volume
|
11.7 gal. (43.5L)
|
Access Port
|
1.25 in. (3.18 cm), removable neoprene plug
|
Shelf Dimensions (W x D)
|
18.5 x 18.5 in. (47 x 47 cm)
|
Internal Dimensions (W x H x D)
|
21.3 x 26.8 x 20 in. (54.1 x 68.1 x 50.8 cm)
|
External Dimensions (W x H x D)
|
26.3 x 39.5 x 25 in. (66.8 x 100.3 x 63.5 cm)
|
Weight
|
265 lb. (120.2 kg.)
|
Electrical Requirements
|
115V, 50/60 Hz, 3.6A, 1Ph
|
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|