|
|
|
|
Popular Alternative Choice
Thermo Forma Steri Cycle 370 CO2 Incubator
- Direct-heat design with high-temperature decontamination.
- In-chamber HEPA filtration achieves Class 100 air quality.
- On-demand 140C sterilization cycle for cleaning.
- Temperature control: ± 0.1°C with range from ambient to 50°C.
Go to Product Details
Fully Recertified!
Thermo Scientific 310 CO2 Incubator
The Thermo Scientific 310 CO2 Incubator is designed to maintain an optimal environment for the incubation of tissue and cell samples. Temperature and Carbon Dioxide levels are maintained and set by the operator, as well as monitoring chamber relative humidity. CO2 is controlled using a thermal conductivity sensor (T/C). The Thermo Scientific Forma 310 CO2 Incubator features temperature control feature from Ambient +5°C to 55°C, and CO2 control from 0 to 20%. This incubator is equipped with an audible and visual alarm system, notifying the user of deviations in the temperature and CO2 levels inside the chamber. This high capacity 6.5 cu. ft. chamber provides maximum culturing space along with an internal access port.
All of our equipment is fully tested to perform
at factory specifications before being shipped.
The Thermo Scientific 310 CO2 Incubator includes:
- > 3 Shelves
- > Humidity Pan
- > New HEPA Filter
- > Calibration and Preventative Maintenance Report
- > Rolling Cart
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific 310 CO2 Incubator Specifications
Temperature
|
Control
|
± 0.1°C Microprocessor PID Control
|
Set point
|
Digital - Touch pad, 0.1°C
|
Range
|
+5°C above ambient to 55°C
|
Uniformity
|
± 0.3°C @ +37°C
|
Tracking Alarm
|
User Programmable (low) indicator
|
Overtemp
|
Tracking, user programmable, action, and indicator
|
Display
|
Digital, LED, 0.1°C increments
|
Temperature Safety
|
Type
|
Extreme temperature safety, action, and indicator
|
Sensor
|
Thermostat, independent of temperature control system
|
Indicator
|
Message center, audible and visual alarms
|
Relative Humidity
|
Control
|
Humidity pan - natural vaporization
|
Humidity with pan
|
95% RH at 37°C
|
Display
|
Optional 1% increments
|
Alarm
|
Low RH with optional RH monitor
|
CO2
|
Control
|
± 0.1% microprocessor PID control
|
Sensor
|
T/C
|
Readability
|
0.1%
|
Range
|
0 to 20%
|
Inlet pressure
|
15psig (1 bar), ± 5 psig (0.3 bar)
|
Display
|
Digital LED, 0.1% increments
|
Thermo Scientific Forma 310 CO2 Incubator Manual
Click Here to View the Technical Specs for the Thermo Scientific Forma 310 CO2 Incubator
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|