Thermo Savant SpeedVac SC110 Concentrator
The Thermo Savant SpeedVac SC110 Concentrator is a state-of-the-art centrifugal vacuum concentrator designed for rapid evaporation of liquids to concentrate or dry various solutes and residues in both aqueous and organic solvents. Its innovative technology includes a vacuum system that connects to an external pump (not included), enhancing solvent evaporation. The concentrator's rotor, while spinning, applies centrifugal force to prevent foaming or bumping, thereby avoiding sample loss, and ensuring optimal analyte recovery. This device is highly versatile, suitable for a range of applications such as concentrating DNA ethanol precipitates, protein hydrolysates, chromatography fractions, and samples for HPLC, GC/MS, immunoassay, and bioassay analysis. It's also used in organic chemistry for residue analysis, including pesticide and drug testing. The SC110 model is equipped with a built-in manual bleeder valve for precise control and features a chamber heater with three temperature settings: Ambient, 43°C, and 65°C, to counteract evaporative cooling and expedite the drying process. The SpeedVac SC110's construction is both durable and chemically resistant, featuring a vacuum chamber made from impregnated aluminum with a fluorocarbon coating and a transparent acrylic cover. It offers a rotor capacity of 64 x 1.5/2.0 ml and includes a 0.5 in. (1.27 cm) O.D. vacuum fitting for the external pump. This concentrator represents a blend of efficiency, reliability, and precision, making it an essential tool in various scientific and research applications. Please Note: An external vacuum pump is NOT INCLUDED and is REQUIRED FOR OPERATION.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Thermo Savant SpeedVac SC110 Concentrator includes:
- > RH64-11 64 x 1.5/2.0 ml Fixed Angle Rotor
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Savant SpeedVac SC110 Concentrator Specifications
Temperature Range
|
Ambient, 43°C, and 65°C setpoints
|
Cover Material
|
Transparent acrylic
|
Vacuum Chamber Material
|
Chemical resistant, impregnated aluminum casting with inert, fluorocarbon coating
|
Vacuum Fitting for External Pump
|
0.5 in. (1.27 cm) O.D.
|
Rotor Capacity
|
64 x 1.5/2.0 ml
|
Dimensions (W x D x H)
|
11 x 18 x 13 in. (29 x 46 x 32 cm)
|
Weight
|
34 lb. (16 kg)
|
Electrical Requirements
|
120V, 60Hz, 6A
|
- Inspect the display panel, replacing any dim or dead pixels to ensure clear visual feedback
- Verify full functionality of all buttons, confirming consistent responsiveness
- Examine the acrylic or glass lid for any cracks or chips, ensuring safety and integrity
- Check the vacuum chamber seal for any tears or cracks that could impact pressure retention
- Test chamber lid functionality, confirming all micro switches report a consistent .04 Ohms
- Verify external connections to the pump or cold trap, ensuring they are properly sealed and leak-free
- Inspect and clean the internal fan and condenser (if applicable) for efficient cooling and airflow
- Examine all internal wiring for frays or breakages, confirming secure and intact connections
- Test motor functionality, replacing motor or bearings if performance is compromised
- Inspect internal pump seals, replacing any worn or damaged seals to maintain vacuum integrity
- Inspect all internal hoses for potential leaks, ensuring tight, clean seals
- Confirm unit is capable of achieving all manufacturer-specified temperature ranges
- Verify the unit can reach specified vacuum levels in accordance with manufacturer standards
- Conduct a 6-8 hour cycle test to confirm consistent operational performance
Technician Inspection and Recertification Checklist: Centrifugal Evaporators
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display panel for dim or dead pixels
|
Completed
|
|
Confirm functionality of all buttons
|
Completed
|
|
Check acrylic/glass lid for cracks or chips
|
Completed
|
|
Inspect vacuum chamber seal for tears or cracks
|
Completed
|
|
Confirm chamber lid micro switches report .04 Ohms
|
Completed
|
|
Verify pump or cold trap connections are sealed and leak-free
|
Completed
|
|
Inspect and clean internal fan and condenser (if applicable)
|
Completed
|
|
Inspect internal wiring for frays or breakages
|
Completed
|
|
Test motor functionality, replace motor/bearings if needed
|
Completed
|
|
Inspect internal pump seals, replace if needed
|
Completed
|
|
Inspect all internal hoses for leaks, ensure clean seals
|
Completed
|
|
Confirm unit achieves specified temperature ranges
|
Completed
|
|
Verify unit reaches specified vacuum levels
|
Completed
|
|
Conduct a 6-8 hour cycle test for operational performance
|
Completed
|
|