Thermo Platform Varimix Rocker Shaker, M79735
The Thermo Scientific Vari Mix Platform Rocker offers angle rocking and variable speed from a gentle motion to a vigorous wave. This platform shaker is a useful tool in many different applications such as hybridization, blotting and staining or destaining gels. This lab rocker has a rocking angle range between 1-24, a speed range between 5 to 30 RPM, and a timer that runs up to 3 hrs. The ample 12 x 10-inch platform or this laboratory rocker has a white silicone nonskid mat to grip your containers, and can support up to 15lbs. All of our equipment will be fully tested to perform at factory specifications before being shipped.
The Thermo Platform Varimix Rocker Shaker, M79735 includes:
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Thermo Platform Varimix Rocker Shaker, M79735 Specifications
Speed Range
|
5 to 30 RPM
|
Platform Angle Variation
|
0 to 24 angle
|
Maximum Load
|
15 lbs.
|
Timer Range
|
0-2hrs (60 Hz), 0-3hrs (50 Hz)
|
Dimensions (W x H x D)
|
12 5/8 x 7 x 12 in
|
Electrical Requirements
|
100-120V, 50/60Hz, 0.1A, 12W
|
Click Here to View all the Technical Specs for the Varimix Rocker Shaker
- Validate that the power supply and PCB deliver stable, correct voltage outputs across all necessary connections, ensuring reliable operation
- Thoroughly inspect the display for any dim, flickering, or dead pixels; replace affected elements to maintain full readability and operational clarity
- Test the full range of button functionality, confirming responsive, accurate input recognition across all settings
- Conduct a detailed inspection of bearings and bearing housing, identifying any worn or damaged bearings, and replace as needed to prevent operational disruption
- Examine the drive belt for signs of cracking, fraying, or material breakdown; replace as required to ensure sustained performance
- Measure and adjust motor drive belt tension to optimal specifications, securing reliable and efficient power transmission
- Verify accuracy of the onboard timer, confirming proper functionality and replacing if it deviates from expected performance
- Calibrate shaker rotation speed with a NIST-traceable tachometer, validating precision within specified operational limits
- Run the shaker continuously for an 8-12 hour testing period, closely observing the shaking platform for smooth, consistent motion, while monitoring for any error indicators such as knocking, skipping, or unusual vibrations
- Perform a comprehensive cosmetic inspection to confirm the device’s exterior meets aesthetic and quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Shakers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed
|
Completed
|
|
Test button functionality for responsiveness and accuracy across all settings
|
Completed
|
|
Inspect bearings and housing for wear; replace as needed to prevent disruption
|
Completed
|
|
Examine drive belt for cracking, fraying, or material breakdown; replace as required
|
Completed
|
|
Measure and adjust motor drive belt tension to optimal specifications
|
Completed
|
|
Verify onboard timer accuracy; replace if it deviates from expected performance
|
Completed
|
|
Calibrate shaker rotation speed using a NIST-traceable tachometer
|
Completed
|
|
Run 8-12 hour continuous test; observe platform for smooth, consistent motion
|
Completed
|
|
Perform comprehensive cosmetic inspection for quality standards compliance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|