Popular Alternative Choice
Benchmark BT3001 Orbi-Shaker Orbital Shaker
- Powerful shaker with 14 lb capacity, speeds up to 300 RPM.
- Standard platform with non-slip surface; MAGic Clamp available.
- Operates between +4C to 65C with a 19mm orbit.
- Compact design: 11 x 13 x 3 in, 120V, 50/60Hz, 600W.
Go to Product Details
No Lead Time and Free Shipping!
Thermo MaxQ 2000 Digital Orbital Shaker
Speed Range 15 to 500 RPM
Operating Temperature 4 to 40 C
- Microprocessor controller ensures accurate time, and speed.
- Advanced internal memory system ensures speed and time set points are maintained during power interruptions.
- Functions effectively in temperatures ranging from 4 to 40°C and humidity conditions of 20 to 80% non-condensing, compatible with incubators, warm rooms, environmental chambers, and refrigerators.
- Equipped with a user-friendly digital control panel for precise settings of speed and time.
- Certified with UL, cUL, and CE standards, ensuring reliability and safety in laboratory settings.
The Thermo MaxQ 2000 Orbital Shaker is ideal for shaking flasks, vials, beakers, Petri dishes, and other sample containers. Variable speed ranges and adjustable timer. Operates in temperature ranges of 4 to 40C and humidity conditions of 20 to 80% non-condensing, enabling them to be used in incubators, warm rooms, environmental chambers and refrigerators. This shaker will be fully tested to perform at factory specifications before being shipped.
The Thermo MaxQ 2000 Digital Orbital Shaker includes:
- > Rubber Mat Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo MaxQ 2000 Digital Orbital Shaker Specifications
Model
|
MaxQ 2000
|
Speed Range
|
15 to 500 rpm
|
Operating Temperature
|
4 to 40 C
|
Orbit
|
1.9cm (0.75 in.)
|
Control
|
Digital
|
Height
|
6.4 in.
|
Hertz
|
50/60Hz
|
Humidity Range
|
20 to 80%
|
Length Exterior
|
17.4 in.
|
Load Bearing Capacity
|
35 lb
|
Motor Type
|
Triple Eccentric drive
|
Shipping Weight
|
47 lb
|
Voltage
|
120V
|
Width Exterior
|
13.5 in.
|
Item Description
|
Small; Digital; 120V 50/60Hz
|
Certifications/Compliance
|
UL, cUL, CE
|
Barnstead MaxQ 2000 Orbital Shaker ManualClick Here to View all the Technical Specs for the MaxQ 2000
- Validate that the power supply and PCB deliver stable, correct voltage outputs across all necessary connections, ensuring reliable operation
- Thoroughly inspect the display for any dim, flickering, or dead pixels; replace affected elements to maintain full readability and operational clarity
- Test the full range of button functionality, confirming responsive, accurate input recognition across all settings
- Conduct a detailed inspection of bearings and bearing housing, identifying any worn or damaged bearings, and replace as needed to prevent operational disruption
- Examine the drive belt for signs of cracking, fraying, or material breakdown; replace as required to ensure sustained performance
- Measure and adjust motor drive belt tension to optimal specifications, securing reliable and efficient power transmission
- Verify accuracy of the onboard timer, confirming proper functionality and replacing if it deviates from expected performance
- Calibrate shaker rotation speed with a NIST-traceable tachometer, validating precision within specified operational limits
- Run the shaker continuously for an 8-12 hour testing period, closely observing the shaking platform for smooth, consistent motion, while monitoring for any error indicators such as knocking, skipping, or unusual vibrations
- Perform a comprehensive cosmetic inspection to confirm the device’s exterior meets aesthetic and quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Shakers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed
|
Completed
|
|
Test button functionality for responsiveness and accuracy across all settings
|
Completed
|
|
Inspect bearings and housing for wear; replace as needed to prevent disruption
|
Completed
|
|
Examine drive belt for cracking, fraying, or material breakdown; replace as required
|
Completed
|
|
Measure and adjust motor drive belt tension to optimal specifications
|
Completed
|
|
Verify onboard timer accuracy; replace if it deviates from expected performance
|
Completed
|
|
Calibrate shaker rotation speed using a NIST-traceable tachometer
|
Completed
|
|
Run 8-12 hour continuous test; observe platform for smooth, consistent motion
|
Completed
|
|
Perform comprehensive cosmetic inspection for quality standards compliance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|