Thermo Labline 4625 Titer Shaker
Shaking Motion 0.120 Inch
Power Requirements 120 VAC, 60 Hz, 120 Watt, 1 Amp
Shaking Speed 40 to 1,100 RPM
- Thermo Labline 4625 Titer Shaker is ideal for many applications involving titer plates.
- The sturdy, heavy-duty design offers reliable performance with an accurate 0.120-inch orbital shaking motion
- A variable speed range of 40 to 1100 rpm produces gentle rotation to vigorous agitation
- Rocker Switch on the control panel allows for selection of continuous shaking or timed cycles up to 5 minutes
- Ribbed platform with non-skid surface and retaining springs offers secure titer plate retention
- Rubber feet for stable placement on countertops prevents the shaker from "walking"
Thermo Labline 4625 Titer Shaker is reliable and dependable combining top-notch and uncompromising quality is ideal for virology, immunology, serology and hematology applications. Two plate retaining springs on platform firmly hold one to four 96-well titer plates. White, non-slip rubber platform mat increases sample visibility. Vortexing motion thoroughly mixes every well evenly. Solid-state speed control provides accurate, continuous regulation of shaking speeds from 40 to 1100 rpm with a 3mm orbit. Built-in electrical timer can be set for continuous operation or timed cycles of up to 5 minutes. Unit shuts off automatically at end of time cycle. Four finger slots provide for easy plate removal. Rugged powder-coated steel body sits on four rubber feet. Equipped with a 10.5x7.5 inch platform. All of our equipment will be fully tested to perform at factory specifications before being shipped.
The Thermo Labline 4625 Titer Shaker includes:
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Labline 4625 Titer Shaker Specifications
Shaking Speed
|
40 to 1,100 RPM
|
Timer
|
Up to 5 minutes in 1 minute increments; continuous
|
Shaking Motion
|
0.12 inch orbit
|
Dimensions (W x D x H)
|
11 x 11 x 7 in (28 x 28 x 18 cm)
|
Electrical Requirements
|
120 VAC, 60 Hz, 120 W, 1 A
|
Thermo Labline 4625 Titer Shaker Manual
Click Here to View all the Technical Specs for the 4625 Titer Shaker
- Validate that the power supply and PCB deliver stable, correct voltage outputs across all necessary connections, ensuring reliable operation
- Thoroughly inspect the display for any dim, flickering, or dead pixels; replace affected elements to maintain full readability and operational clarity
- Test the full range of button functionality, confirming responsive, accurate input recognition across all settings
- Conduct a detailed inspection of bearings and bearing housing, identifying any worn or damaged bearings, and replace as needed to prevent operational disruption
- Examine the drive belt for signs of cracking, fraying, or material breakdown; replace as required to ensure sustained performance
- Measure and adjust motor drive belt tension to optimal specifications, securing reliable and efficient power transmission
- Verify accuracy of the onboard timer, confirming proper functionality and replacing if it deviates from expected performance
- Calibrate shaker rotation speed with a NIST-traceable tachometer, validating precision within specified operational limits
- Run the shaker continuously for an 8-12 hour testing period, closely observing the shaking platform for smooth, consistent motion, while monitoring for any error indicators such as knocking, skipping, or unusual vibrations
- Perform a comprehensive cosmetic inspection to confirm the device’s exterior meets aesthetic and quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Shakers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed
|
Completed
|
|
Test button functionality for responsiveness and accuracy across all settings
|
Completed
|
|
Inspect bearings and housing for wear; replace as needed to prevent disruption
|
Completed
|
|
Examine drive belt for cracking, fraying, or material breakdown; replace as required
|
Completed
|
|
Measure and adjust motor drive belt tension to optimal specifications
|
Completed
|
|
Verify onboard timer accuracy; replace if it deviates from expected performance
|
Completed
|
|
Calibrate shaker rotation speed using a NIST-traceable tachometer
|
Completed
|
|
Run 8-12 hour continuous test; observe platform for smooth, consistent motion
|
Completed
|
|
Perform comprehensive cosmetic inspection for quality standards compliance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|