Thermo Forma 3120 CO2 Water Jacketed Incubator
The Thermo Forma 3120 Water-Jacketed CO2 Incubator combines precise CO2 control. Our equipment is fully tested to perform at factory specifications before being shipped.
The Thermo Forma 3120 CO2 Water Jacketed Incubator includes:
- > Calibration and Preventative Maintenance Report
- > 3 Shelves
- > Humidity Pan
- > New Hepa Filter
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Forma 3120 CO2 Water Jacketed Incubator Specifications
Temperature Range
|
+5C above ambient to 55C
|
Temperature Uniformity
|
0.2C @ 37C
|
Humidity System
|
3 L pan
|
Humidity Range
|
Ambient to 95% @ 37C
|
Interior Volume
|
6.5 cu ft (184.1 L)
|
Thermo Forma 3120 Manual
Click Here to View all the Technical Specs for the 3120
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|