Thermo Forma 3110 CO2 Water Jacketed Incubator
Interior Volume 6.5 cu ft.
Temperature Range
Ambient +5°C to 55°C
- Precise CO2 control and temperature for reliable incubation of critical cell cultures.
- Highly popular and trusted incubator in the market.
- Forma Series II Water Jacketed CO2 Incubator with precise CO2 control, elevated RH, and thermal conductivity sensor.
- Outstanding temperature stability and superior parameter recovery characteristics.
- Innovative continuous contamination control technology for a clean environment.
- 100% HEPA filtration for achieving Class 100 air quality by entrapping particulate contaminants.
- Ensures temperature retention and quick recovery from ambient temperature swings and power variations.
- Includes calibration and preventative maintenance report, shelves, humidity pan, HEPA filter, power cord, operator's manual, and a 180-Day Parts and Labor Warranty.
The Thermo Forma 3110 Water-Jacketed CO2 Incubator combines
precise CO2 control and temperature and is the most popular incubator on the market.
The Forma Series II Water Jacketed CO2
Incubator, with its precise CO2 control, elevated RH, and a thermal conductivity sensor, is the first choice for many
researchers. These incubators feature unsurpassed temperature
stability and superior parameter recovery characteristics, with innovative
continuous contamination control technology.
The Forma 3110's 100% HEPA filtration quickly achieves Class 100 air quality. The HEPA filter entraps
particulate air contaminants and prevents their escape, and filter efficiency
increases as particulates and micro-organisms are entrapped.
Temperature retention and quick recovery are especially important for
researchers working with critical cell cultures. The Thermo Forma 3110 water
jacketed incubator provides maximum thermal protection and quick recovery from
swings in ambient temperature and power variations.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The Thermo Forma 3110 CO2 Water Jacketed Incubator includes:
- > Calibration and Preventative Maintenance Report
- > 3 Shelves
- > Humidity Pan
- > New Hepa Filter
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Forma 3110 Water-Jacketed CO2 Incubator Specifications
Temperature:
|
|
Control
|
± 0.1°C
|
Temp Range
|
Ambient +5°C to 55°C
|
|
|
Tracking Alarm
|
Programmable High/Low
|
CO2
|
|
CO2 Control
|
Better than 0.1%
|
CO2 Range
|
0-20%
|
Inlet Pressure
|
15 PSIG (1.0 bar)
|
CO2 Sensor
|
T/C or IR
|
Readability
|
0.1%
|
Tracking Alarm
|
User programmable High/Low
|
Humidity:
|
|
RH
|
Ambient or 95% @ 37°C
|
Humidity Pan
|
3.2qt standard
|
Optional Display
|
In 1% increments
|
Inlet Pressure:
|
|
Maximun Per Channel
|
40.8 sq ft.
|
Standard Maximum
|
4,17
|
Safety:
|
|
Sensor
|
Precision thermistor
|
Controller
|
Independent Analog
|
Settings
|
0.1C
|
Construction:
|
|
Water Jacket Volume
|
12.7 gallons
|
Interior Volume
|
6.5 cu ft.
|
Interior
|
stainless steel
|
Exterior
|
coiled rolled steel
|
Outlet Door Gasket
|
Four sided magnetic vinyl
|
Interior Door Gasket
|
feather edge silicon.
|
Thermo Forma 3110 CO2 Water Jacketed Incubator Manual
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|