MJ Research PTC-100 Thermal Cycler
Thermal Range 0°C to 105°C
Thermal Uniformity ± 0.4°C well-to-well
Ramping Rate Up to 1.2°C/sec
- Compatible with both 96-well and 60-well microplates for versatile sample processing
- Offers a wide temperature range from 0°C to 105°C with high accuracy (± 0.3°C)
- Features a heated lid for oil-free operation and easy pressure and height adjustment
- Designed for durability, capable of enduring up to 10,000 temperature cycles
- Allows indefinite material storage at any temperature for accurate incubation rates
- Stores approximately 320 programs with 14 pre-programmed protocols included
- Equipped with power-failure protection, preserving data for up to twelve hours after a power interruption
- Flexible ramping rates with up to 1°C/sec for 60-well block and 1.2°C/sec for 96-well block
The MJ Research PTC-100 Thermal Cyler is a robust
thermocycler that operates with
96-well and 60-well microplates. This thermocycler is
reliable and easy-to-use, while also offering various temperature
configurations from 0°C to 105°C with temp accuracy of ± 0.3°C. The PTC-100 has
a heated lid for oil-free operation, which allows for convenient pressure and
height adjustment via a thumbwheel. Crafted to withstand repeated cycling, the
PTC-100 thermocycler can follow any temperature sequence up to 10,000 times and
can store material indefinitely at any temperature, allowing for highly
accurate incubation rates. The MJ Research PTC-100 can store approximately 320
programs and comes with 14 pre-programmed protocols. Due to the PTC's
power-failure protection feature, data is safe after power interruption for up
to twelve hours.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The MJ Research PTC-100 Thermal Cycler includes:
- > 96 Well Alpha Block
- > NIST Traceable Calibration Certificate
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
MJ Research PTC-100 Thermal Cycler Specifications
Thermal Range
|
0°C - 105°C
|
Accuracy
|
± 0.5°C of programmed target at 60°C
|
Thermal Uniformity
|
± 0.4°C well-to-well within 30 sec of arrival at 60°C
|
Ramping rate
|
60-well block: up to 1°C/sec
96-well: up to 1.2°C/sec
|
Sample Capacity
|
60-well block: 60 x 0.5 ml tubes
96-well block: 96 x 0.2 ml tubes or
one 96-well plate
|
Line Voltage
|
100-240 VAC rms (no adjustment needed among voltages within these
ranges)
|
PTC-100 manual
We also carry the PTC-150, PTC-200, PTC-225, and PTC-240. Click Here to View all the Technical Specs for the PTC-100
MJ Research PTC-100 Thermal Cycler
- Validate that the power supply and PCB consistently deliver accurate, stable voltages across all circuits and connections, confirming operational reliability
- Thoroughly inspect the display for any dim, flickering, or dead pixels, replacing compromised segments to ensure clear visibility and precise data display
- Verify complete functionality and responsiveness of all buttons and knobs, confirming accurate and reliable input across all controls
- Ensure the machine completes all internal diagnostic checks with zero errors, indicating system readiness and functional integrity
- Confirm the machine accurately identifies attached blocks (96-well, dual 48, etc.), with seamless recognition across different configurations
- Conduct a detailed inspection of internal wiring and all electrical connections, checking for any signs of breaks, disconnections, or tears; replace or repair any underperforming components
- Examine the connector between the block and base for bent or broken pins, replacing as needed to maintain secure, reliable connections
- Calibrate well temperatures with a NIST-traceable thermometer, verifying accuracy and uniformity across all wells to meet specification standards
- Perform a comprehensive cosmetic inspection, ensuring all exterior surfaces and components are free of damage, wear, or defects, maintaining aesthetic and professional quality
Technician Inspection and Recertification Checklist: Thermal Cycler
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, accurate voltage outputs from power supply and PCB across circuits
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed for clarity
|
Completed
|
|
Verify functionality and responsiveness of all buttons and knobs
|
Completed
|
|
Confirm machine completes internal diagnostics with zero errors
|
Completed
|
|
Ensure accurate block identification across configurations (96-well, dual 48, etc.)
|
Completed
|
|
Inspect internal wiring and connections for integrity; repair or replace as needed
|
Completed
|
|
Check connector between block and base for bent or broken pins; replace as necessary
|
Completed
|
|
Calibrate well temperatures with a NIST-traceable thermometer for accuracy
|
Completed
|
|
Conduct cosmetic inspection to ensure quality and professionalism of all surfaces
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|