Stirling Ultracold SU780XLE -86C Upright Ultra-Low Freezer
The Stirling Ultracold SU780XLE -86C Upright Ultra-Low Freezer is at the forefront of ultra-low temperature (ULT) freezer technology, offering an unparalleled temperature range of -20°C to -86°C, adjustable in 1°C increments. This freezer stands out with its industry-leading steady-state temperature variation of ± 1°C over time and an impressive 35-minute door opening recovery time to -80°C within ± 1°C of the setpoint, as per the ENERGY STAR Final Test Method. It is equipped with connectivity options for Building Management Systems (BMS)/Building Automation Systems (BAS) or third-party monitoring systems, alongside a real-time temperature display for constant monitoring. With a voluminous 780L capacity in a compact design, the SU780XLE offers the largest storage capacity per square foot of floor space in its class, accepting universal power sources for versatile installation. Its modulated cooling capacity and 100% adaptive control ensure faster temperature pull-down and recovery times, while the advanced Stirling engine technology employs only two moving parts, eliminating the need for compressors and significantly reducing the risk of failure. The SU780XLE is not only a powerhouse in protecting research materials but also champions environmental sustainability. It consumes up to 70% less energy than traditional compressor-based units and 45% less than similar-sized ENERGY STAR rated freezers, utilizing EPA SNAP-approved 100% natural refrigerants and environmentally friendly manufacturing processes. This results in a significantly smaller carbon footprint. Economically, the freezer is designed to protect your operating budget by offering up to 70% reduction in electric utility costs compared to legacy freezers, alongside minimized heat output and HVAC operation costs. It reduces the need for extensive floor space, infrastructure, and backup power, all while maintaining the lowest ongoing maintenance and service costs. With its manual defrost method, helium-charged free-piston Stirling cooling engine, and gravity-driven thermosiphon heat transport system, the SU780XLE is a smart choice for research organizations seeking reliability, efficiency, and environmental responsibility.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Stirling Ultracold SU780XLE -86C Upright Ultra-Low Freezer includes:
- > 5 Shelves
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Stirling Ultracold SU780XLE -86C Upright Ultra-Low Freezer Specifications
Temperature Range
|
-20°C to -86°C, adjustable in 1°C increments
|
Volume
|
780L (27.5 cu. ft.)
|
Cooling Engine
|
Helium charged free-piston Stirling engine with continuous modulation
|
Heat Transport System
|
Gravity driven thermosiphon
|
Refrigerant
|
R-170 (Ethane) 90 grams
|
Evaporator
|
Cold wall (inner liner)
|
Heat Rejection
|
Finned heat exchanger with forced air cooling
Air inlet: above freezer door, below mechanical compartment
Air outlet: right side of top cover, upward
|
Defrost Method
|
Manual
|
Controller Type
|
Microprocessor with touchscreen input and display
|
Warm and Cold Alarms
|
Fully adjustable
|
Temperature Log
|
30 days available graphically
|
Battery Back-up
|
12-hour control battery back-up for touchscreen
|
Energy Use
|
6.67 kWh/day at -75°C
|
Pull Down from 25°C Ambient
|
6.5 hours to -80°C (empty cabinet)
|
Recovery from Door Opening
|
35 minutes at -80°C
|
Warm-up Profile
|
2.5 hours to -60°C at -80°C (empty cabinet)
6.5 hours to -40°C at -80°C (empty cabinet)
|
Heat Dissipation
|
981 BTU/hr (load to HVAC) at -80°C (empty cabinet)
|
Interior Dimensions (H x D x W)
|
60.7 x 27.8 x 29.1 in. (154.2 x 70.5 x 74 cm)
|
Exterior Dimensions (H x D x W)
|
78.5 x 34.3 x 36 in. (199.4 x 87 x 91.5 cm)
|
Weight
|
655 lb. (297 kg.)
|
Electrical Requirements
|
120-240V, 50/60Hz, 1200W
|
- Inspect the display for pixel clarity and brightness, replacing any dim or dead pixels.
- Test all buttons for responsiveness, resistance, and durability; replace any that don’t meet standards.
- Check the door seal and lock for airtightness, replacing any compromised components.
- Run diagnostics on PCB test points for voltage and current accuracy, replacing faulty components.
- Verify compressor output and airflow, ensuring all internal fans function properly.
- Cross-check internal temperature probe accuracy with a NIST-traceable probe, recalibrating if needed.
- Inspect wiring for frays or tears, replacing compromised sections as necessary.
- Test the start capacitor's capacitance and replace if it’s outside the specified range.
- Replace all internal filters and thoroughly clean condenser fins for efficient heat dissipation.
- Verify the alarm battery holds a full charge and replace if there is any degradation.
- Operate the unit for 72 hours to confirm stable temperatures and operational consistency.
- Perform a final cosmetic check, repairing or touching up surfaces as needed for a professional appearance.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Inspect display for pixel issues and replace if needed | Completed | |
Test button functionality and replace any defective buttons | Completed | |
Examine door seal and lock functionality; replace if compromised | Completed | |
Conduct PCB test point diagnostics and replace as needed | Completed | |
Verify compressor and internal fan functionality | Completed | |
Confirm temperature probe accuracy with NIST-traceable probe | Completed | |
Inspect all wiring for frays, tears, or breakages | Completed | |
Test start capacitor capacitance range; replace if needed | Completed | |
Replace all internal filters and clean condenser fins | Completed | |
Test alarm battery for full functionality | Completed | |
Run 72-hour test with data recorder to confirm stable temperatures | Completed | |
Conduct cosmetic inspection and address any issues | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|