Sorvall RC5C Plus Refrigerated Centrifuge
The Sorvall RC5C Plus Refrigerated Floor Centrifuge offers a user-friendly interface for large-capacity efficient centrifugation. The refrigeration system on the RC5C Plus offers temperatures from -20C to +40C with an automatic temperature compensation system and auto slow start and stop for rate control. Reach speeds up to 21,000 RPM with maximum RCF of 51,427 x g. This high-performance centrifuge combines the flexibility of microprocessor control with the durability, reliability, and convenience of a heavy-duty, direct-drive motor system. All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped. Please confirm this rotor meets your application needs and request a different rotor, if necessary. Available rotors can be found in our Rotor Compatibility Guide.
The Sorvall RC5C Plus Refrigerated Centrifuge includes:
- > SS34 Rotor
- > Calibration Report
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Sorvall RC5C Plus Refrigerated Centrifuge Specifications
Speed Selection Range (rpm)
|
50 to 21,000 RPM
|
Speed Control Accuracy
|
1% or 100 rpm, whichever is greater
|
Maximum Relative Centrifugal Force
|
51,427 g
|
Temperature Selection Range
|
-20C to +40C
|
Mass (Weight)
|
362 kilograms, 800 pounds
|
Noise Level
|
<62dB
|
Heat Output
|
4 kW (13,500 BTU)
|
Electrical Requirements
|
208 V, 60 Hz
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly
Sorvall RC5C Plus Operator's Manual
Click Here to View all the Technical Specs for the RC5C Plus
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task |
Completed? |
Remarks? |
Confirm power supply output voltages match specified requirements |
Completed |
|
Measure and verify voltages at PCB test points for accuracy |
Completed |
|
Confirm rotor compatibility and detection for proper identification |
Completed |
|
Inspect exterior for cosmetic or structural damage |
Completed |
|
Check door mechanism alignment for secure opening and closing |
Completed |
|
Inspect and replace gas strut with OEM part as needed |
Completed |
|
Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
Replace and confirm functionality of all lid lock switches |
Completed |
|
Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
Test compressor (if applicable) for proper operation and cooling |
Completed |
|
Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
Check rotor motor for stability, noise, and integrity |
Completed |
|
Inspect and clear condensation drain to prevent buildup |
Completed |
|
Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|