Scientific Industries SI-3000 MultiMagstir Genie Magnetic Stirrer
The Scientific Industries SI-3000 MultiMagstir Genie Magnetic Stirrer is a four position magnetic stirrer that provides low to high speed stirring and editable memory profiles. Microprocessor controls offer precision and reproducibility. Boasts a large capacity that can hold up to 4- 1L flasks/beakers or 3- 1L spinner flasks and a maximum sample weight of 10.0 Kg. The manual mode provides stir, stir reverse, and time functions.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Scientific Industries SI-3000 MultiMagstir Genie Magnetic Stirrer includes:
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Scientific Industries SI-3000 MultiMagstir Genie Magnetic Stirrer Specifications
Stirrer Speed
|
10 to 1,200 RPM
|
Platform Size
|
14 x 12.5 in (35.6 x 31.8 cm)
|
Electrical Requirements
|
120V, 50/60Hz, 0.6A
|
Timer
|
1 sec to 99 hrs
|
Power Levels
|
Low, Medium, High
|
Ramp Rates
|
5 to 300 RPM/sec
|
Stirring Positions
|
4
|
Maximum Capacity
|
1L per position
|
Weight
|
7.1 lbs
|
Markings
|
ETL Listed / CE
|
- Inspect the display for clarity and brightness, replacing any dim or dead pixels to ensure optimal performance.
- Examine external regulators for structural integrity and damage, replacing if necessary (if applicable).
- Assess the functionality of temperature, CO₂, and H₂O sensors, verifying accuracy and consistency (if applicable).
- Check all internal wiring for frays, tears, or breakages, addressing any compromised sections.
- Confirm software settings are properly configured to match the specifications of internal components.
- Inspect and clean the fan for efficient airflow, replacing if wear or malfunction is detected.
- Verify the condition of wheels and leveling mounts, ensuring stability and ease of movement.
- Inspect door gaskets and latches for sealing integrity, replacing if compromised to maintain proper enclosure.
- Calibrate temperature accuracy to 37°C using an NIST-traceable temperature probe for precision.
- Calibrate CO₂ levels to 5% using a Fyrite combustion test kit to verify environmental control (if applicable).
- Confirm temperature and CO₂ stability over a 24-hour period to ensure consistent conditions.
- Conduct a cosmetic inspection, verifying the unit meets appearance and cleanliness standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Inspect display for clarity and replace if needed | Completed | |
Examine external regulators for integrity and damage | Completed | |
Assess temperature, CO₂, and H₂O sensors for accuracy | Completed | |
Check internal wiring for frays, tears, or breakages | Completed | |
Confirm software settings match internal component specs | Completed | |
Inspect and clean fan, replacing if wear is detected | Completed | |
Verify condition of wheels and leveling mounts | Completed | |
Inspect door gaskets and latches for sealing integrity | Completed | |
Calibrate temperature to 37°C with NIST-traceable probe | Completed | |
Calibrate CO₂ to 5% with Fyrite combustion test kit | Completed | |
Confirm 24-hour temperature and CO₂ stability | Completed | |
Conduct cosmetic inspection to ensure appearance standards | Completed | |