Scientific Industries SI-1500 Orbital-Genie Orbital Shaker
The Scientific Industries SI-1500 Orbital Shaker with Ratcheting Clamps is a heavy duty orbital shaker with flexibility for a variety of vessels on a single platform. The SI-1500 offers whisper quiet operation with fully programmable options. Control speed from 10 RPM to 300 RPM with fully programmable time, reversing time, acceleration and deceleration, orbit reversing, etc. Customizable mixing action allows for anything from a smooth vortex to sloshing aeration. Programmable ramping allows for gentle stirring increase to preserve samples and avoid spills.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Scientific Industries SI-1500 Orbital-Genie Orbital Shaker includes:
- > Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Scientific Industries SI-1500 Orbital-Genie Orbital Shaker Specifications
Speed
|
10 300 RPM including stir reverse
|
Timer
|
1 min - 99 hours w/Alarm or Continuous
|
Orbit
|
35mm
|
Selectable Ramp Rates
|
10 - 100 rpm/sec
|
Product Weight
|
21 Kg (46 lbs.)
|
Dimensions (base)
|
460 x 460mm (18 x 18in)
|
Overall (D x W x H)
|
590 x 500mm (23 x 20in)
|
Electrical Requirements
|
120V, 2A
|
Click Here to View all the Technical Specs for the Orbital-Genie
- Validate that the power supply and PCB deliver stable, correct voltage outputs across all necessary connections, ensuring reliable operation
- Thoroughly inspect the display for any dim, flickering, or dead pixels; replace affected elements to maintain full readability and operational clarity
- Test the full range of button functionality, confirming responsive, accurate input recognition across all settings
- Conduct a detailed inspection of bearings and bearing housing, identifying any worn or damaged bearings, and replace as needed to prevent operational disruption
- Examine the drive belt for signs of cracking, fraying, or material breakdown; replace as required to ensure sustained performance
- Measure and adjust motor drive belt tension to optimal specifications, securing reliable and efficient power transmission
- Verify accuracy of the onboard timer, confirming proper functionality and replacing if it deviates from expected performance
- Calibrate shaker rotation speed with a NIST-traceable tachometer, validating precision within specified operational limits
- Run the shaker continuously for an 8-12 hour testing period, closely observing the shaking platform for smooth, consistent motion, while monitoring for any error indicators such as knocking, skipping, or unusual vibrations
- Perform a comprehensive cosmetic inspection to confirm the device’s exterior meets aesthetic and quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Shakers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed
|
Completed
|
|
Test button functionality for responsiveness and accuracy across all settings
|
Completed
|
|
Inspect bearings and housing for wear; replace as needed to prevent disruption
|
Completed
|
|
Examine drive belt for cracking, fraying, or material breakdown; replace as required
|
Completed
|
|
Measure and adjust motor drive belt tension to optimal specifications
|
Completed
|
|
Verify onboard timer accuracy; replace if it deviates from expected performance
|
Completed
|
|
Calibrate shaker rotation speed using a NIST-traceable tachometer
|
Completed
|
|
Run 8-12 hour continuous test; observe platform for smooth, consistent motion
|
Completed
|
|
Perform comprehensive cosmetic inspection for quality standards compliance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|