Sartorius Practum 1102-1S Precision Balance
The Sartorius Practum 1102-1S Precision Balance produces consistent readings and excellent repeatability ensured by Sartorius quality and technology developed and designed in Germany. Features include: h
igh reliability and durability provided by mechanical overload protection, supervisor lock for preventing unintentional changes to the setup settings, and effortless weighing thanks to the ergonomic and sturdy draft shield.
Included with the Sartorius Practum 1102-1S Precision Balance:
Includes:
> Power Cord > Full 180 Day Parts and Labor Warranty
Sartorius Practum 1102-1S Precision Balance Specifications
Max. Capacity | 1100 g | Readability | 10 mg | Repeatability (sd) | 10 mg | Linearity | 30 mg | Calibration Type | Manual | RS-232 | Yes | Stabilization Time | 3 Seconds | Electrical Requirements | 115V, 50/60Hz | Languages | English, French, German, Hungarian, Italian, Polish, Portuguese, Russian, Spanish, Turkish, Chinese, Japanese, Korean | Weighing Units | g, ct, mom, kg, lb, oz, ozt, tlh, tlt, GN, dwt, mg, parts lb, tlc, Kt, tol, bat, MS, N | Weigh Pan Size (Diameter) | 180 mm | Dimensions (D x W x H) | 360 x 216 x 320 mm (14.1 x 8.5 x 12.6 in) |
Sartorius Practum 1102-1S Precision Balance Manual
- Verify power supply output for precise voltage delivery in accordance with equipment specifications
- Assess full functionality of display and control buttons, replacing any defective or deteriorating components as necessary
- Confirm balance is properly leveled to ensure measurement accuracy and stability
- Examine windscreen glass for cracks or chips, replacing if structural integrity is compromised
- Inspect all internal wiring and electrical connections for secure attachment and check for any loose connections or damaged connectors
- Check power cord integrity, ensuring there are no frays, cuts, or damage affecting function or safety
- Execute internal or external calibration to confirm accuracy of weighing measurements
- Review cosmetic condition, confirming no visible damage or wear that affects functionality
- Verify successful taring of the weighing pan to zero out prior measurements accurately
- Ensure RS232/printer communication functions correctly, validating connection and data transmission (if applicable)
Technician Inspection and Recertification Checklist: Balances
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Verify power supply output voltage
|
Completed
|
|
Assess display and control button functionality, replacing faulty components
|
Completed
|
|
Confirm balance leveling for measurement accuracy
|
Completed
|
|
Inspect windscreen glass for cracks, replace if necessary
|
Completed
|
|
Check internal wiring and electrical connections for security
|
Completed
|
|
Inspect power cord for frays or damage
|
Completed
|
|
Execute internal/external calibration for measurement accuracy
|
Completed
|
|
Review cosmetic condition for visible damage
|
Completed
|
|
Verify weighing pan tares successfully
|
Completed
|
|
Ensure RS232/printer communication is functional (if applicable)
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|