Sanyo MCO-19AIC(UV) CO2 Incubator Dual Stack
Temperature Range Ambient +5°C to 50°C
- Advanced humidity and temperature control technology for precise culture conditions
- Direct heat and air jacket system eliminates the need for water and provides even heat distribution
- Dual detector infrared CO2 sensor maintains constant conditions regardless of humidity or temperature changes
- Safe Cell UV contamination control system ensures a sterile environment
- Spacious interior with 170L volume and two stainless steel trays
- Rigid polyurethane foamed-in-place insulation for efficient temperature control
- PID control system for temperature and CO2 levels with digital displays
- Air circulation with fan assistance for uniform conditions
The Sanyo MCO-19AIC(UV) CO2 Incubator features unprecedented humidity and temperature control technology, which provides both flexible and robust stability control for cultures. The new direct heat and air jacket system eliminates the inconvenient use of water in the incubator and provides heat from the door, side, and bottom of the unit. The Sanyo MCO-19AIC utilizes a dual detector infrared CO2 sensor, which is not affected by changes in humidity or temperature, maintains constant conditions via the microprocessor control, and automatically calibrates using room air temperature as reference. Also features Safe Cell UV contamination control system.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Sanyo MCO-19AIC(UV) CO2 Incubator Dual Stack includes:
- > 2 x Shelves Per Unit
- > 2 x Humidity Pans
- > Stacking Kit
- > Calibration and Preventative Maintenance Report
- > 2 x Power Cords
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Sanyo MCO-19AIC(UV) CO2 Incubator Dual Stack Specifications
Temperature Range
|
Ambient +5°C to 50°C
|
Temperature Uniformity
|
± 0.25°C at 37°C
|
Temperature Stability
|
± 0.1°C at 37°C
|
CO2 Range
|
1 to 20%
|
CO2 Control Deviation
|
± 0.15%
|
CO2 Sensor Type
|
IR
|
Required Gas Pressure
|
0.03 MPaG (0.3 kgf/cm2 G, 4.3 psiG)
|
Relative Humidity
|
95% ± 5%
|
Volume
|
6 cu ft. (170L) [Per Unit]
|
Air Filter
|
0.3 um, efficiency: 99.97% or higher
|
UV Lamp
|
4W X 1 (ozone-free emission)
|
Interior Material
|
Stainless steel containing copper
|
Access Port Diameter
|
30 mm
|
Alarms
|
Automatic set temperature alarm, automatic set CO2 density alarm, upper limit temperature alarm, door alarm, low humidifying water alarm, UV lamp failure alarm
|
Maximum Heat Emission
|
1,120 kJ/h
|
Shelf Dimensions (W x D x H)
|
17 x 17 x 4.7 in. (45 x 45 x 12 cm)
|
Internal Dimensions (W x D x H)
|
19 x 20.6 x 26.2 in. (49 x 52.3 x 66.5 cm) [Per Unit]
|
External Dimensions (W x D x H)
|
24.4 x 27.9 x 35.4 in. (62 x 71 x 90 cm) [Per Unit]
|
Weight
|
205 lb. (93 kg.) [Per Unit]
|
Electrical Requirements
|
110-120V, 60 Hz, 2.8A, 350W [Per Unit]
|
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|