Revco ULT3030A18 -30C Laboratory Freezer
The Revco ULT3030A18 -30C Freezer features a microprocessor controlled system, gravity air circulation, industrial quality cabinet construction, and extra-strength refrigeration compressors selected for industrial, clinical and scientific use. This ULT model has audible/visual warnings for over temperature, under temperature, and power failure, with visual status reports on critical functions.
Graphic thermometer confirms proper operation at-a-glance, shows normal, high or low temperature condition; steady green for normal, flashing bulb for under temperature.
Enlarged digital display, resolution to within 0.1C, improves visibility. Alarm silence, ringback, and automatic reset functions. Automatic defrost and condensate removal with forced-air circulation for fast recovery. With improved temperature uniformity, the ULT3030A can provide faster recovery after door is opened. All of our equipment is fully tested to perform at factory specifications before being shipped.
Included with the Revco ULT3030A18 -30C Laboratory Freezer: Includes: > 4 Shelves > Power Cord > Operator's Manual > Full 180 Day Parts and Labor Warranty Revco ULT3030A18 -30C Laboratory Freezer Specifications
Temperature
|
-30C
|
Capacity
|
29.2 cu ft (826 L)
|
Exterior Dimensions (L x W x H)
|
37.2 x 34 x 79.2 in (94.5 x 86.4 x 201.2 cm)
|
Interior Dimensions (L x W x H)
|
29 x 30 x 58 in (73.7 x 76.2 x 147.3 cm)
|
Electrical Requirements
|
115V, 60Hz, 16A
|
Revco ULT3030A18 Freezer Manual
Click Here to View the Technical Specs for the Revco ULT3030A18 Freezer
- Inspect the display for pixel clarity and brightness, replacing any dim or dead pixels.
- Test all buttons for responsiveness, resistance, and durability; replace any that don’t meet standards.
- Check the door seal and lock for airtightness, replacing any compromised components.
- Run diagnostics on PCB test points for voltage and current accuracy, replacing faulty components.
- Verify compressor output and airflow, ensuring all internal fans function properly.
- Cross-check internal temperature probe accuracy with a NIST-traceable probe, recalibrating if needed.
- Inspect wiring for frays or tears, replacing compromised sections as necessary.
- Test the start capacitor's capacitance and replace if it’s outside the specified range.
- Replace all internal filters and thoroughly clean condenser fins for efficient heat dissipation.
- Verify the alarm battery holds a full charge and replace if there is any degradation.
- Operate the unit for 72 hours to confirm stable temperatures and operational consistency.
- Perform a final cosmetic check, repairing or touching up surfaces as needed for a professional appearance.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Inspect display for pixel issues and replace if needed | Completed | |
Test button functionality and replace any defective buttons | Completed | |
Examine door seal and lock functionality; replace if compromised | Completed | |
Conduct PCB test point diagnostics and replace as needed | Completed | |
Verify compressor and internal fan functionality | Completed | |
Confirm temperature probe accuracy with NIST-traceable probe | Completed | |
Inspect all wiring for frays, tears, or breakages | Completed | |
Test start capacitor capacitance range; replace if needed | Completed | |
Replace all internal filters and clean condenser fins | Completed | |
Test alarm battery for full functionality | Completed | |
Run 72-hour test with data recorder to confirm stable temperatures | Completed | |
Conduct cosmetic inspection and address any issues | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|