Popular Alternative Choice
Scilogex iPette Plus Electronic Pipettor, 0.5-10µl
- Scilogex iPette Plus Electronic Pipettor, 0.5-10µl: Reduces user strain with ergonomic design.
- Features: Clear LED display, tip ejection, use in narrow tubes, 8hr lithium-ion battery.
- Included: AC Adapter, USB Cable, Stand, Operator's Guide, Documentation, 1yr Warranty.
- Specifications: Precision 3.0/0.5%, Increment 0.01µl, Volume 0.5-10µl, Accuracy 3.5/1.0%.
Go to Product Details
Free Shipping!
Rainin EDP3-Plus Electronic Pipette Set
The Rainin EDP3-Plus Electronic Pipette Set is a variable volume air displacement pipette set for aspirating and dispensing volumes between 20
l and 1200l, with precision, accuracy, and ease. The EDP3-Plus pipettes feature a large LCD display screen, along with a user-friendly keypad for easy volume adjustment and mode selection. These pipettes can be operated in a number of different functions, including pipet mode, multidispense mode, dilute mode, manual mode, and titrate mode. Also included is the E3-RCS Rapid Charging Station, which charges three EDP3-Plus pipettes in sequence. To extend battery life, the EDP3-Plus display "sleeps" after ten minutes of inactivity and will wake with the press of any key or the trigger. Ergonomics are also taken into account with these pipettes, featuring a finger hook and trigger system for comfortable pipetting.
All of our equipment will be fully tested to perform at factory specifications before being shipped.
The Rainin EDP3-Plus Electronic Pipette Set includes:
- > Rainin EDP3-Plus 200ul Single Channel Electronic Pipette
- > Rainin EDP3-Plus 200ul 8-Channel Electronic Pipette
- > Rainin EDP3-Plus 1200ul 8-Channel Electronic Pipette
- > Charging Station
- > Full 180 Day Parts and Labor Warranty
Rainin EDP3-Plus Electronic Pipette Set Specifications
Volume
|
Increment
|
Accuracy
|
Precision
|
20l
100l
200l
20l
100l
200l
100l
600l
1200l
|
0.2l
0.2l
1.0l
|
2.5%
0.8%
0.8%
2.5%
0.8%
0.8%
3.6%
0.8%
0.8%
|
1.00%
0.25%
0.15%
1.00%
0.25%
0.15%
0.60%
0.20%
0.15%
|
Click Here to View the Technical Specs for the Rainin EDP3-Plus Electronic Pipette Set
- Thoroughly inspect and clean all internal components, identifying and replacing any damaged or compromised parts
- Replace all essential wear components, including O-rings, seals, tip holders, and friction rings, to ensure optimal performance
- Examine plunger and seal holders for signs of rust, wear, or damage, replacing components as necessary for operational integrity
- Verify that no leakage is present by aspirating a sample and observing for any fluid loss
- Confirm that the Mettler Toledo AB104-S balance is precisely calibrated for accurate pipette calibration measurements
- Perform a detailed gravimetric calibration, recording all findings in the comprehensive calibration report
- Ensure that all external parts are meticulously cleaned and free from any foreign contaminants
- Test the tip ejector for smooth and reliable operation, addressing any misalignment or obstruction
- Affix a dated calibration sticker to indicate recent and verified calibration
- Conduct a detailed cosmetic inspection, confirming that all surfaces and finishes meet established quality and aesthetic standards
Technician Inspection and Recertification Checklist: Pipettes
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Thoroughly inspect and clean all internal components; replace damaged parts
|
Completed
|
|
Replace essential wear components like O-rings, seals, and friction rings
|
Completed
|
|
Examine plunger and seal holders for wear or damage; replace if needed
|
Completed
|
|
Verify no leakage is present by aspirating sample and checking for fluid loss
|
Completed
|
|
Confirm Mettler Toledo AB104-S balance calibration for accurate measurements
|
Completed
|
|
Perform detailed gravimetric calibration and record findings in report
|
Completed
|
|
Clean all external parts meticulously to remove any contaminants
|
Completed
|
|
Test tip ejector for smooth, reliable operation; correct misalignments
|
Completed
|
|
Affix dated calibration sticker to indicate verified calibration
|
Completed
|
|
Conduct cosmetic inspection to ensure quality of surfaces and finishes
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|