Ohaus Voyager Analytical Balance, Model V02140
All of our equipment is fully tested to perform at factory specifications before being shipped.
Included with the Ohaus Voyager Analytical Balance, Model V02140:
Includes: > Power Cord > Calibration Report > Operator's Guide > Full 180 Day Parts and Labor Warranty
Ohaus Voyager Analytical Balance Specifications
Max. Capacity | 210 g | Readability | 0.1 mg | Repeatability (sd) | 0.1 mg | Linearity | 0.2 mg | Calibration Type | Internal | RS-232 | Yes | Stabilization Time | ? 4 Seconds | Electrical Requirements | 120V, 50/60Hz | Weighing Modes | g, mg, oz, oz t, ct, dwt, taels (3), mommes, gn, ti, N, custom unit | Functions | Parts counting, differential weighing, quick check, statistics, formulation, filling, animal weighing, percent formulation, check weighing, density determination, pipette calibration | Sensitivity Drift PPM/C | 3 | Operating Temperature Range | With Internal Calibration: 10C to 40C Without Internal Calibration: 10C to 30C | Dimensions (W x D x H) | 9 x 15.25 x 14 in | Ohaus Voyager Analytical Balance Manual
Click Here to View all the Technical Specs for the Ohaus Voyager
- Verify power supply output for precise voltage delivery in accordance with equipment specifications
- Assess full functionality of display and control buttons, replacing any defective or deteriorating components as necessary
- Confirm balance is properly leveled to ensure measurement accuracy and stability
- Examine windscreen glass for cracks or chips, replacing if structural integrity is compromised
- Inspect all internal wiring and electrical connections for secure attachment and check for any loose connections or damaged connectors
- Check power cord integrity, ensuring there are no frays, cuts, or damage affecting function or safety
- Execute internal or external calibration to confirm accuracy of weighing measurements
- Review cosmetic condition, confirming no visible damage or wear that affects functionality
- Verify successful taring of the weighing pan to zero out prior measurements accurately
- Ensure RS232/printer communication functions correctly, validating connection and data transmission (if applicable)
Technician Inspection and Recertification Checklist: Balances
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Verify power supply output voltage
|
Completed
|
|
Assess display and control button functionality, replacing faulty components
|
Completed
|
|
Confirm balance leveling for measurement accuracy
|
Completed
|
|
Inspect windscreen glass for cracks, replace if necessary
|
Completed
|
|
Check internal wiring and electrical connections for security
|
Completed
|
|
Inspect power cord for frays or damage
|
Completed
|
|
Execute internal/external calibration for measurement accuracy
|
Completed
|
|
Review cosmetic condition for visible damage
|
Completed
|
|
Verify weighing pan tares successfully
|
Completed
|
|
Ensure RS232/printer communication is functional (if applicable)
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|