New Brunswick S41i CO2 Incubator Shaker w/ Sterilization Cycle
The New Brunswick S41i CO2 Incubator Shaker w/ Sterilization Cycle combines the best of two worlds: a highly innovative shaker and the functionality of a CO2 incubator into one unit. This CO2 Incubator Shaker features a large, easy-to-read touchscreen display for quick viewing/setting of multiple parameters, help screens, trend graphs, diagnostics, and more. The S41i precisely controls temperature from Ambient +4°C to 50°C, shaking speed from 25 to 400 RPM, and CO2 from 0.2 to 20%, providing the ideal environment for growing mammalian and stem cells in culture.
The New Brunswick S41i CO2 Incubator Shaker w/ Sterilization Cycle utilizes the renowned triple eccentric drive which provides stable, uniform and vibration-free motion connected to the shaking platform by four encapsulated support rods. The drive mechanism located below the chamber ensures that heat generated by the drive does not affect the critical environment within the chamber. The large 24 in. x 14 in. platform accommodates flasks up to 4 liters providing flexibility with your cultures. This incubator shakers also features a fully automatic High Temperature Disinfection cycle which heats the internal chamber to 120°C, maintains the temperature for 3.5 hours, and then allows the chamber to cool down to the programmed temperature.
The New Brunswick S41i CO2 Incubator Shaker w/ Sterilization Cycle includes:
- > 24" x 14" Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
New Brunswick S41i CO2 Incubator Shaker w/ Sterilization Cycle Specifications
Speed Range
|
25 to 400 RPM
|
Speed Control Accuracy
|
± 1%
|
Temperature Range
|
Ambient +4ºC to 50ºC
|
Temperature Uniformity
|
± 0.25ºC at 37°C
|
Temperature Stability
|
± 0.1°C at 37°C
|
CO2 Range
|
0.2 to 20%
|
CO2 Uniformity
|
± 0.1%
|
CO2 Stability
|
± 0.2% at 5% CO2
|
Required Gas Pressure
|
0.05 MPa (0.5 bar / 7.2 psi)
|
Display
|
Touchscreen
|
Timer
|
Continuous
|
Orbit Size
|
2.5 cm (1 in.)
|
Platform Size
|
24 x 14 in. (61.2 x 35.6 cm)
|
Chamber Dimensions (W x D x H)
|
27.2 x 17.5 x 21.2 in. (69.1 x 44.4 x 53.8 cm)
|
Overall Dimensions (W x D x H)
|
34.4 x 28.7 x 33.5 in. (87.5 x 73 x 85 cm)
|
Weight
|
340 lb. (192 kg.)
|
Electrical Requirements
|
120V, 60Hz
|
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Examine chamber glass for cracks or chips, replace as needed | Completed | |
Inspect door seals for wear, replace to ensure tight seal | Completed | |
Check display brightness, replace dim or dead pixels | Completed | |
Test all buttons for full functionality and responsiveness | Completed | |
Replace platform assembly bearings for reliable motion | Completed | |
Inspect drive belt for wear, replace if torn or cracked | Completed | |
Examine wiring for frays or breakages, repair as needed | Completed | |
Verify PCB test points, replace faulty components | Completed | |
Check drive belt tension for proper alignment | Completed | |
Confirm compressor functionality, clean filter and fins | Completed | |
Calibrate internal temperature with NIST probe | Completed | |
Calibrate platform RPM with NIST-traceable tachometer | Completed | |
Conduct cosmetic inspection for appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|