New Brunswick Innova 44 Incubator Shaker Dual Stack
Temperature Range Ambient +5°C to 80°C
Temperature Accuracy
± 0.1C from 30C to 40C
± 0.5C over the remaining range
- Triple-eccentric counterbalanced drive for high-speed and heavy workload support
- Solid-state, brushless DC motor with self-calibrating mechanism
- Proportional/Integral microprocessor control for precise shaking speed regulation
- Audible and visual alarms for enhanced safety and monitoring
- Temperature range from ambient +5°C to 80°C with accuracy of ± 0.1C from 30C to 40C
- Low watt-density resistance heaters with high-temperature safety thermostat
- Large 30 x 18 in platforms for accommodating large samples
- Built-in timer adjustable from 0.01 to 99.9 hours with continuous operation option
The New Brunswick Innova 44 Incubator Shaker Dual Stack is a cutting-edge instrument engineered to support a wide range of high-demand laboratory applications with superior precision and reliability. Featuring a triple-eccentric counterbalanced drive, this dual stack system handles heavy workloads with ease while its solid-state, brushless DC motor with a self-calibrating mechanism ensures smooth and consistent operation. Proportional/Integral microprocessor control delivers precise shaking speed regulation across its operational range, and both audible and visual alarms enhance safety and monitoring throughout each run. With a temperature operating range from ambient +5°C to 80°C and exceptional uniformity of ±0.25°C at 37°C maintained by low watt-density resistance heaters and a high-temperature safety thermostat, this incubator shaker guarantees optimal environmental conditions for sample processing. A large 30 x 18 inch platform accommodates extensive sample sizes, and a built-in timer adjustable from 0.01 to 99.9 hours—including a continuous operation option—provides the flexibility required for diverse experimental protocols.
All of our equipment will be fully tested to perform at factory specifications before being shipped.
The New Brunswick Innova 44 Incubator Shaker Dual Stack includes:
- > 18 x 30 in. Platform [Per Unit]
- > Stacking Kit
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
New Brunswick Innova 44 Incubator Shaker Dual Stack Specifications
Speed Range
|
25 to 300 RPM (stacked)
|
Speed Accuracy
|
± 1 RPM
|
Temperature Range
|
Ambient +5C to 80C
|
Temperature Accuracy
|
± 0.1C from 30C to 40C
±0.5C over the remaining range
|
Timer
|
0.01 to 99.9 hours, or continuous
|
Platform Size
|
18 x 30 in.
|
Orbit
|
1 in. (2.5 cm)
|
Chamber Dimensions (W x D x H)
|
36 x 26 x 18 in. (91 x 66 x 46 cm) [Per Unit]
|
Overall Dimensions (W x D x H)
|
53.1 x 33.2 x 52 in. (134.9 x 84.3 x 132 cm)
|
Weight
|
525 lb. (238 kg) [Per Unit]
|
Electrical Requirements
|
120V, 60Hz, 11A, 1100W [Per Unit]
|
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Examine chamber glass for cracks or chips, replace as needed | Completed | |
Inspect door seals for wear, replace to ensure tight seal | Completed | |
Check display brightness, replace dim or dead pixels | Completed | |
Test all buttons for full functionality and responsiveness | Completed | |
Replace platform assembly bearings for reliable motion | Completed | |
Inspect drive belt for wear, replace if torn or cracked | Completed | |
Examine wiring for frays or breakages, repair as needed | Completed | |
Verify PCB test points, replace faulty components | Completed | |
Check drive belt tension for proper alignment | Completed | |
Confirm compressor functionality, clean filter and fins | Completed | |
Calibrate internal temperature with NIST probe | Completed | |
Calibrate platform RPM with NIST-traceable tachometer | Completed | |
Conduct cosmetic inspection for appearance standards | Completed | |