New Brunswick Innova 42R Refrigerated Incubator Shaker
The New Brunswick Innova 42R Refrigerated Incubator Shaker is a high-capacity, space-efficient lab device that operates within a speed range of 25 to 400 RPM and temperature range of Ambient -20° to 80°C. Designed for ergonomic and safe use, the Innova 42R features top-mounted controls and display, eliminating the need for bending down and preventing accidental parameter changes. It is easy to clean, thanks to a spill diverter and a built-in, drainable spill pan with rounded corners, which also serves as a water reservoir to reduce sample evaporation. The unit offers lab position flexibility, being stackable and suitable for placement on or under benches. It includes a large viewing window and internal light for easy sample monitoring without opening the door.
This refrigerated incubator shaker is reliable for 24/7 operation even under high loads, thanks to its proven Innova triple eccentric counterbalanced drive, supporting a maximum load of 15 kg. It offers programmability with continuous, timed, or multi-step programming (15 steps), and can store up to four programs. The Innova 42R features a speed control accuracy of ± 1 RPM and speed display in 1 RPM increments. It also provides a temperature uniformity of ± 0.25°C at 37°C and an accuracy of ± 0.1°C at 37°C. Additional features include a 0.1 to 99.9 hours timer or continuous operation, a 1 in. (2.5 cm) orbit, and alarms for deviations in speed and temperature, as well as expired timers. The device's triple-eccentric counterbalanced drive is complemented by nine permanently lubricated ball bearings, and it uses a long-life, low-watt density resistance-type heater with a high-temperature thermostat. The platform measures 18 x 18 in. (46 x 46 cm), further enhancing its utility in various laboratory settings.
All of our equipment will be fully tested to perform at factory specifications before being shipped.
The New Brunswick Innova 42R Refrigerated Incubator Shaker includes:
- > 18 x 18 in. Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
New Brunswick Innova 42R Refrigerated Incubator Shaker Specifications
Speed Range
|
25 to 400 RPM
|
Speed Control Accuracy
|
± 1 RPM
|
Speed Display
|
1 RPM Increments
|
Temperature Range
|
Ambient -20°C to 80°C (Minimum of 4°C)
|
Temperature Uniformity
|
± 0.25°C at 37°C
|
Temperature Accuracy
|
± 0.1°C at 37°C, ± 0.5°C over the remaining range
|
Timer
|
0.1 to 99.9 hours, or continuous
|
Orbit
|
1 in. (2.5 cm)
|
Alarms
|
Visible and audible warning indication when speed deviates more than 5 rpm or temperature more than 1 °C from setpoints, and when timer has expired.
|
Drive Mechanism
|
Triple-eccentric counterbalanced drive with nine permanently lubricated ball bearings
|
Heaters
|
Long-life, low-watt density resistance-type heater with high temperature thermostat
|
Platform Size (W x D)
|
18 x 18 in. (46 x 46 cm)
|
Chamber Dimensions (W x D x H)
|
20.4 x 20.3 x 18.8 in. (51.7 x 51.5 x 47.7 cm)
|
Overall Dimensions (W x D x H)
|
25 x 29.3 x 32.2 in. (63.5 x 74.5 x 81.8 cm)
|
Weight
|
266 lb. (121 kg)
|
Electrical Requirements
|
120V, 60Hz, 1Ph
|
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Examine chamber glass for cracks or chips, replace as needed | Completed | |
Inspect door seals for wear, replace to ensure tight seal | Completed | |
Check display brightness, replace dim or dead pixels | Completed | |
Test all buttons for full functionality and responsiveness | Completed | |
Replace platform assembly bearings for reliable motion | Completed | |
Inspect drive belt for wear, replace if torn or cracked | Completed | |
Examine wiring for frays or breakages, repair as needed | Completed | |
Verify PCB test points, replace faulty components | Completed | |
Check drive belt tension for proper alignment | Completed | |
Confirm compressor functionality, clean filter and fins | Completed | |
Calibrate internal temperature with NIST probe | Completed | |
Calibrate platform RPM with NIST-traceable tachometer | Completed | |
Conduct cosmetic inspection for appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|