New Brunswick Innova 4230 Refrigerated Incubator Shaker
Speed Range 25 to 400 RPM
Temperature Range Ambient -20°C to 80°C
Temperature Uniformity ± 0.3°C
- Heavy-duty stainless steel construction with a large double-paned glass window for internal sample viewing
- Equipped with an adjustable stainless steel shelf and a 13-watt fluorescent light for enhanced visibility
- Features an internal AC receptacle specifically for powering internal devices like magnetic stirrers or shakers
- Built-in shaker with a triple eccentric counterbalanced drive, offering a shaking range of 25-400 rpm (non-stacked) and 25-300 rpm (stacked)
- Temperature range from 20°C below ambient, with a minimum of 4°C to a maximum of 75°C (50 Hz units) or 80°C (60 Hz units)
- Audible and visual alarms for temperature deviations greater than 1.0°C, with an option to deactivate the audible alarm
- Robust refrigeration system using 1/5 HP CFC-Free (R134A) Refrigerant, ensuring efficient cooling
- Large internal dimensions (20 1/2"W x 20 1/2"L x 24.8"H) with a total weight of 300 lbs, making it suitable for substantial laboratory applications
The New Brunswick Innova 4230 Refrigerated Incubator Shaker
is the perfect refrigerated incubator shaker for almost all lab situations. With
a heavy gauge stainless steel frame, both refrigeration and heating
capabilities, and a powerful internal shaking platform, this Refrigerated
incubator is fully capable of handling just about any sample parameters.
The internal Cabinet of the 4230 is composed of sturdy heavy
gauge stainless steel and has a double paned glass window on the outside to
allow internal sample viewing. The cabinet contains two adjustable stainless
steel shelves and a fluorescent light for improved visibility. The 4230 also includes an internal AC
receptacle for the sole purpose of the powering of internal devices.
The built in shaker features an audible and visible alarm,
control, display, timer and other onboard installations. The internal shaker
drive is a triple eccentric counter balanced drive with 9 sealed for life ball
bearings and has a motion of 3/4" circular standard. The rpm range is between 25
and 400 non stacked, and 25-300 stacked. The onboard timer has a time range of
up to 99.9 hours settable in 0.1 hours.
The temperature of the 4230 ranges from 20°C below ambient
with a minimum of 4°C to maximum of 75°C for 50 Hz units or 80°C for 60 Hz
units. Temperature alarms provide indication when temperature deviates > 1.0°C
All of our equipment is fully tested to perform at factory specifications before being shipped.
The New Brunswick Innova 4230 Refrigerated Incubator Shaker includes:
- > Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
New Brunswick Innova 4230 Refrigerated Incubator Shaker Specifications
Cabinet
|
Construction
|
Heavy gauge stainless steel
|
Door
|
26 3/8"H x 22"W (670 mm x 559 mm)
|
Double Paned Glass Window
|
19 1/4"H x 11 3/8"W (489 mm x 289 mm)
|
Fluorescent Light
|
13 watt - provides illumination inside chamber
|
Shelves
|
2 Stainless Steel 20"W x 20"L (508 mm x 508 mm)-adjustable- (1/2" (12.7
mm) centers)
|
Internal AC Receptacle
|
200W AC-waterproof w/ mating connector-can be used w/ magnetic stirrer,
spinner flask, rocker, roller & shakers
|
Dimensions
|
Platform
|
18"W x 18"L (457 mm x 457 mm)
|
Overall
|
24.4"W x 28.7"L x 36.4"H (620 mm x 729 mm x 925 mm)
|
Chamber
|
20 1/2"W x 20 1/2"L x 24.8"H (520 mm x 520 mm x 630 mm)
|
Electrical
|
1500 Watts, 120/100/200/240V, 50/60 Hz
|
Indicator
|
3 digit LED
|
Refrigeration
|
1/5 HP uses CFC-Free (R134A) Refrigerant
|
Recorder Output
|
Two 0-5 Volt-tracks temperature & speed w/ optional recorder
|
Weight
|
300 lbs
|
Shaking
|
Alarm
|
Audible & visible indicates when speed deviates ± 5 rpm, audible
alarm can be deactivated
|
Control
|
± 1 rpm, acceleration control minimizes splashing caused by sudden
starts and stops
|
Display
|
1 rpm increments
|
Drive
|
Triple-eccentric counterbalanced drive w/ 9 sealed for life, ball
bearings
|
Motion
|
3/4"(19 mm) circular, standard 1?(25.4) optional
|
Motor
|
1/15 HP, 3-phase solid-state, brushless ball bearing DC motor
|
Range
|
25 to 400 rpm, 25 to 300 rpm for stacked unit
|
Hour Timer
|
Programmable shaking continuous - 99.9 hours-setable in 0.1 hr.
increments-alarm sounds at end of shaking period
|
Temperature
|
Alarms
|
Audible & Visual-when temperature deviates > 1.0°C-audible
alarm can be deactivated
|
Ambient
|
15-35°C/Humidity-up to 90% non-condensing
|
Control
|
± 0.1°C-alternation between two setpoints programmable via electronic
timer
|
Heaters
|
Resistance w/ high temperature safety cutoff
|
Range
|
20°C below ambient with a minimum of 4°C to maximum of 75°C for 50 Hz
units or 80°C for 60 Hz units
|
Refrigeration
|
1/5 HP/CFC-Free (R134A) Refrigerant
|
Uniformity
|
± .3C NOM
|
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Examine chamber glass for cracks or chips, replace as needed | Completed | |
Inspect door seals for wear, replace to ensure tight seal | Completed | |
Check display brightness, replace dim or dead pixels | Completed | |
Test all buttons for full functionality and responsiveness | Completed | |
Replace platform assembly bearings for reliable motion | Completed | |
Inspect drive belt for wear, replace if torn or cracked | Completed | |
Examine wiring for frays or breakages, repair as needed | Completed | |
Verify PCB test points, replace faulty components | Completed | |
Check drive belt tension for proper alignment | Completed | |
Confirm compressor functionality, clean filter and fins | Completed | |
Calibrate internal temperature with NIST probe | Completed | |
Calibrate platform RPM with NIST-traceable tachometer | Completed | |
Conduct cosmetic inspection for appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|