New Brunswick Innova 2350 Orbital Shaker
The New Brunswick Scientific Innova 2350 Orbital Shaker is a benchtop open air platform shaker for use with interchangeable platforms, clamps, and test tubes. The Innova 2350 is a dependable shaker for ambient conditions on the bench, in an incubator, or in a warm or cold room. The shaker's triple-eccentric counterbalanced drive in cast iron housing provides vibration and trouble free operation for years. All of our equipment is fully tested to perform at factory specifications before being shipped.
The New Brunswick Innova 2350 Orbital Shaker includes:
- > 24" x 36" Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
New Brunswick Innova 2350 Orbital Shaker Specifications
Speed Control
|
± 1 rpm
|
Timer
|
0.1 - 99.9 h
|
Power Supply
|
100/120/220/240 V, 50/60 Hz
|
Dimensions (W x D x H)
|
91.4 x 61 x 16.3 cm (36 x 24 x 6.4 in)
|
Balanced Drive Mechanism
|
Triple-eccentric with permanently lubricated ball bearings
|
Motor Type
|
Solid state, DC brushless motor
|
Orbit
|
2.5 cm (1 in)
|
Platform Size
|
91 x 61 cm (36 x 24 in)
|
Speed Range
|
25 - 300 rpm
|
New Brunswick Innova 2350 ManualClick Here to View all the Technical Specs for the Innova 2350
New Brunswick Innova 2300 Shaker
- Validate that the power supply and PCB deliver stable, correct voltage outputs across all necessary connections, ensuring reliable operation
- Thoroughly inspect the display for any dim, flickering, or dead pixels; replace affected elements to maintain full readability and operational clarity
- Test the full range of button functionality, confirming responsive, accurate input recognition across all settings
- Conduct a detailed inspection of bearings and bearing housing, identifying any worn or damaged bearings, and replace as needed to prevent operational disruption
- Examine the drive belt for signs of cracking, fraying, or material breakdown; replace as required to ensure sustained performance
- Measure and adjust motor drive belt tension to optimal specifications, securing reliable and efficient power transmission
- Verify accuracy of the onboard timer, confirming proper functionality and replacing if it deviates from expected performance
- Calibrate shaker rotation speed with a NIST-traceable tachometer, validating precision within specified operational limits
- Run the shaker continuously for an 8-12 hour testing period, closely observing the shaking platform for smooth, consistent motion, while monitoring for any error indicators such as knocking, skipping, or unusual vibrations
- Perform a comprehensive cosmetic inspection to confirm the device’s exterior meets aesthetic and quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Shakers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed
|
Completed
|
|
Test button functionality for responsiveness and accuracy across all settings
|
Completed
|
|
Inspect bearings and housing for wear; replace as needed to prevent disruption
|
Completed
|
|
Examine drive belt for cracking, fraying, or material breakdown; replace as required
|
Completed
|
|
Measure and adjust motor drive belt tension to optimal specifications
|
Completed
|
|
Verify onboard timer accuracy; replace if it deviates from expected performance
|
Completed
|
|
Calibrate shaker rotation speed using a NIST-traceable tachometer
|
Completed
|
|
Run 8-12 hour continuous test; observe platform for smooth, consistent motion
|
Completed
|
|
Perform comprehensive cosmetic inspection for quality standards compliance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|