|
|
|
|
New Brunswick Innova 2100 Shaker, M1194-000
The New Brunswick Scientific Innova 2100 Platform Shaker is in excellent condition and comes with a recertified platform. The Innova 2100 microprocessor controls allow for uniform motion and accurate regulation of shaking speed.
Audible and visual alarms come standard with built-in timer. Shaking speeds from 25 to 500 RPM with 1 RPM speed control.
All of our equipment will be fully tested to perform at factory specifications before being shipped.
The New Brunswick Innova 2100 Shaker, M1194-000 includes:
- > Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
New Brunswick Innova 2100 Shaker, M1194-000 Specifications
Speed
Range
|
25
to 500 RPM
|
Speed
Control
|
1 RPM
|
Timer
|
0.1
to 99.9 hours
|
Balanced
Drive Mechanism
|
Triple-eccentric
with permanently lubricated ball bearings
|
Motor
Type
|
Solid
state, DC brushless motor
|
Orbit
|
1.9
cm (0.75 in)
|
Power
Supply
|
120V,
50/60Hz
|
New Brunswick Innova 2100 Manual Click Here to View all the Technical Specs for the Innova 2100
- Validate that the power supply and PCB deliver stable, correct voltage outputs across all necessary connections, ensuring reliable operation
- Thoroughly inspect the display for any dim, flickering, or dead pixels; replace affected elements to maintain full readability and operational clarity
- Test the full range of button functionality, confirming responsive, accurate input recognition across all settings
- Conduct a detailed inspection of bearings and bearing housing, identifying any worn or damaged bearings, and replace as needed to prevent operational disruption
- Examine the drive belt for signs of cracking, fraying, or material breakdown; replace as required to ensure sustained performance
- Measure and adjust motor drive belt tension to optimal specifications, securing reliable and efficient power transmission
- Verify accuracy of the onboard timer, confirming proper functionality and replacing if it deviates from expected performance
- Calibrate shaker rotation speed with a NIST-traceable tachometer, validating precision within specified operational limits
- Run the shaker continuously for an 8-12 hour testing period, closely observing the shaking platform for smooth, consistent motion, while monitoring for any error indicators such as knocking, skipping, or unusual vibrations
- Perform a comprehensive cosmetic inspection to confirm the device’s exterior meets aesthetic and quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Shakers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed
|
Completed
|
|
Test button functionality for responsiveness and accuracy across all settings
|
Completed
|
|
Inspect bearings and housing for wear; replace as needed to prevent disruption
|
Completed
|
|
Examine drive belt for cracking, fraying, or material breakdown; replace as required
|
Completed
|
|
Measure and adjust motor drive belt tension to optimal specifications
|
Completed
|
|
Verify onboard timer accuracy; replace if it deviates from expected performance
|
Completed
|
|
Calibrate shaker rotation speed using a NIST-traceable tachometer
|
Completed
|
|
Run 8-12 hour continuous test; observe platform for smooth, consistent motion
|
Completed
|
|
Perform comprehensive cosmetic inspection for quality standards compliance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|