New Brunswick I26R Refrigerated Incubator Shaker
Speed Range 25 to 400 RPM (25 to 250 RPM for Stacked Units)
Temperature Range Ambient -15C to 60C
- Refrigerated temperature range from ambient -15°C to 60°C with ±0.25°C uniformity at 37°C
- Speed range of 25 to 400 RPM (25 to 250 RPM for stacked units) with ±1 RPM accuracy
- Robust triple-eccentric counterbalanced drive ensures vibration-free operation
- Microprocessor controls with large LED display for precise regulation of speed, temperature, and time
- 30 x 18-inch slide-out platform for easy access to all samples
- Stackable design allows up to three units to maximize capacity
- Supports constant speed and temperature or timed shaking modes
- Durable cast iron housing for long-lasting, trouble-free operation
The New Brunswick I26R Refrigerated Incubator Shaker combines precise temperature control with advanced shaking capabilities, making it an ideal choice for demanding laboratory applications. Building on the features of the standard I26 model, the I26R adds powerful refrigeration capabilities, offering a temperature range of ambient -15°C to 60°C with exceptional uniformity of ±0.25°C at 37°C. This stackable orbital shaker features a robust triple-eccentric counterbalanced drive mechanism housed in durable cast iron, ensuring vibration-free operation even under high-speed applications and heavy workloads.
The I26R provides a speed range of 25 to 400 RPM (25 to 250 RPM for stacked units) and an orbit size of 2.5 cm (1 in). Its microprocessor controls allow precise regulation of temperature, speed, and running time, while the user-friendly keypad and large LED display make it easy to set and monitor parameters. The spacious 30 x 18-inch slide-out platform ensures convenient access to samples, including those in the back, while the ability to stack up to three units maximizes capacity without increasing the footprint. With a focus on durability, efficiency, and flexibility, the I26R is designed to deliver consistent performance for years to come.
The New Brunswick I26R Refrigerated Incubator Shaker includes:
- > 30" x 18" Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
New Brunswick I26R Refrigerated Incubator Shaker Specifications
Speed Range
|
25 to 400 RPM (25 to 250 RPM for Stack of 2 to 3 Units)
|
Speed Control Accuracy
|
± 1 RPM
|
Temperature Range
|
Ambient -15°C to 60°C
|
Temperature Uniformity
|
± 0.25°C at 37C
|
Timer
|
0.1 to 99.9 hours, or continuous
|
Program Modes
|
Constant speed and temperature and timed shaking
|
Orbit Size
|
2.5 cm (1 in)
|
Platform Size
|
30 x 18 in (76 x 46 cm)
|
Chamber Dimensions (W x D x H)
|
34 x 23 x 15.5 in (86.4 x 58.4 x 39.4 cm)
|
Overall Dimensions (W x D x H)
|
50 x 30.26 x 27.56 in (127 x 77 x 70 cm)
|
Weight
|
370 lb. (167.8 kg)
|
Electrical Requirements
|
120V, 60Hz, 1Ph, 1500VA
|
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Examine chamber glass for cracks or chips, replace as needed | Completed | |
Inspect door seals for wear, replace to ensure tight seal | Completed | |
Check display brightness, replace dim or dead pixels | Completed | |
Test all buttons for full functionality and responsiveness | Completed | |
Replace platform assembly bearings for reliable motion | Completed | |
Inspect drive belt for wear, replace if torn or cracked | Completed | |
Examine wiring for frays or breakages, repair as needed | Completed | |
Verify PCB test points, replace faulty components | Completed | |
Check drive belt tension for proper alignment | Completed | |
Confirm compressor functionality, clean filter and fins | Completed | |
Calibrate internal temperature with NIST probe | Completed | |
Calibrate platform RPM with NIST-traceable tachometer | Completed | |
Conduct cosmetic inspection for appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|