New Brunswick Galaxy 170R CO2 Incubator
Temperature Range 4°C above ambient to 50°C
Interior Dimensions
21.2 x 27.2 x 17.5 in.
- Revolutionary control system that allows for easy programming, control, and monitoring of the chamber conditions.
- Direct heating system encompasses the interior of the incubator to provide even temperature conditions within the chamber.
- Infrared Sensor ensures reliable control and monitoring of CO2 levels.
- Seamless chamber provides easy access for cleaning and maintenance.
- Sealed inner glass door with a cam action lock to allow viewing of the cultures without compromising the internal atmosphere.
The Galaxy 170 R CO2 incubator is designed to provide stable and homogeneous atmosphere required for cell culture by controlling temperature, carbon dioxide and optionally oxygen for cultivating samples and cells from biological laboratories. The fan-less design offers more usable space and eliminates the fan, reducing contamination. A six-sided direct heating system provides a uniform incubated environment to gently bathe cells, and the deep-drawn, stainless steel chamber eliminates seams or welds and allows easy cleaning. Other features include an LCD display with display readout of temperature and CO2 level, sealed inner glass door, reliable IR CO2 sensor, RS-232 communication port, and alarm settings.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The New Brunswick Galaxy 170R CO2 Incubator includes:
- > Calibration Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
New Brunswick Galaxy 170R CO2 Incubator Specifications
Exterior Dimensions (W x H x D)
|
27.9 x 33.3 x 27.6 in
|
Interior Dimensions (W x H x D)
|
21.2 x 27.2 x 17.5 in
|
Volume
|
165 L
|
Temperature Range
|
4 C above ambient temperature to 50 C
|
Temperature Control Increment
|
0.1 C
|
Temperature Stability
|
0.1 C at 37 C
|
Temperature Uniformity
|
0.3 C at ambient 20 C 25 C
|
CO2 Range
|
0.2 - 20 %
|
CO2 Control Increment
|
0.1 %
|
CO2 Stability
|
0.2 % at 5 % CO2
|
CO2 Uniformity
|
0.1 %
|
Gas Connections
|
Inner diameter of 6.5 mm and outer diameter of 10 mm
|
Required Gas Pressure
|
0.05 MPa (0.5 bar/7.2 psi)
|
Reservoir Capacity
|
2.5 L
|
Humidity Control
|
95 % at 37 C
|
Electrical
|
110 V 127 V 10 %, 50 Hz 60 Hz, 1000W
|
Benchmark ST-45 CO2 Incubator, 45L, 115V w/ Two Shelves
- Temperature Range Ambient +5 to 60C
- Includes Two Shelves
- Brand New
- 2 Year Manufacturer's Warranty
Go to Product Details
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|