New Brunswick G-25 Incubator Shaker
Speed Range 40 to 400 RPM
Temperature Range Ambient +5°C to 60°C
Platform Size 18 x 30 in. (45.7 x 76.2 cm)
- Combines controlled temperature incubation with continuous-duty shaking for versatile applications
- Durable triple eccentric drive ensures reliable and reproducible agitation
- High-performance ball bearing motor minimizes friction and stress for long-term operation
- Hermetically sealed, double-pane viewing window allows for easy observation without heat loss
- Adjustable air intake vent enables room air circulation without disrupting temperature control
- Thoroughly insulated growth chamber for consistent temperature regulation
- Robust construction with a weight of 325 lbs, ensuring stability during operation
- Compact design with chamber dimensions of 22 x 34.5 x 12.5 in, ideal for lab settings
The New Brunswick G-25 Incubator Shaker combines the advantages of a controlled temperature incubator with the efficiency of a continuous duty shaking apparatus. The G-25 is built for continuous service and reproducible agitation. Acting through a triple eccentric drive, a powerful ball bearing motor rotates the shaking platform horizontally in a 1 in. (2.5 cm) circular orbit with a speed range of 40 to 400 RPM. To minimize friction and stress, the platform rotates on nine life-time ball bearings of high thrust and load bearing capacity.
The environment growth chamber of the New Brunswick G-25 is thoroughly insulated and has an extremely low heat transfer coefficient for close temperature uniformity across the range of Ambient +5°C to 60°C. Work can be observed through a hermetically sealed, double pane viewing window in the incubator lid. An adjustable air intake vent in the chamber permits room air to circulate through the incubator when required, without affecting temperature regulation.
The New Brunswick G-25 Incubator Shaker includes:
- > 18" x 30" Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
New Brunswick G-25 Incubator Shaker Specifications
Speed Range
|
40 to 400 RPM
|
Temperature Range
|
Ambient +5°C to 60°C
|
Orbit Size
|
2.5 cm (1 in.)
|
Platform Size
|
18 x 30 in. (45.7 x 76.2 cm)
|
Chamber Dimensions (L x W x H)
|
22 x 34.5 x 12.5 in. (55.9 x 87.6 x 31.8 cm)
|
Overall Dimensions (L x W x H)
|
22 x 45.5 x 30.5 in. (55.9 x 115.6 x 77.5 cm)
|
Weight
|
325 lb. (147.4 kg.)
|
Electrical Requirements
|
120V, 50/60Hz, 1Ph
|
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Examine chamber glass for cracks or chips, replace as needed | Completed | |
Inspect door seals for wear, replace to ensure tight seal | Completed | |
Check display brightness, replace dim or dead pixels | Completed | |
Test all buttons for full functionality and responsiveness | Completed | |
Replace platform assembly bearings for reliable motion | Completed | |
Inspect drive belt for wear, replace if torn or cracked | Completed | |
Examine wiring for frays or breakages, repair as needed | Completed | |
Verify PCB test points, replace faulty components | Completed | |
Check drive belt tension for proper alignment | Completed | |
Confirm compressor functionality, clean filter and fins | Completed | |
Calibrate internal temperature with NIST probe | Completed | |
Calibrate platform RPM with NIST-traceable tachometer | Completed | |
Conduct cosmetic inspection for appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|