Molecular Devices SpectraMax M2 Multilabel Microplate Reader
Microplate Formats 6-,12-,24-,48-,96-, 384-wells
Absorbance Wavelength Range 200 to 1000 nm
Fluorescence Wavelength Range (Excitation/Emission) 250 to 850 nm/250 to 850 nm
- Dual-monochromator, multi-detection microplate reader with dual-mode cuvette port
- Compatible with 6-,12-,24-,48-,96-, 384-well microplates, suitable for various assay formats
- Features absorbance (UV-Vis Abs) and fluorescence intensity (FI) detection modalities
- Offers endpoint, kinetic, spectrum, and area-well scanning read types with PathCheck technology
- Allows simultaneous absorbance and fluorescence intensity assays with a single read command
- Supports a wide absorbance wavelength range from 200 to 1000 nm for versatile applications
- Equipped with a Xenon flash lamp and photomultiplier tube for high-quality light detection
- Temperature control from Ambient +4°C to 45°C along with plate shaking capability, ensuring optimal conditions for various assays
The SpectraMax M2 is a dual-monochromator, multi-detection
microplate reader with a dual-mode cuvette port and 96 or 384 microplate reading
capability. Detection modalities include absorbance (UV-Vis Abs) and
fluorescence intensity (FI) and have optical performance comparable to a
top-of-the-line dedicated spectrofluorometer or spectrophotometer. Endpoint,
kinetic, spectrum and area-well scanning read types and PathCheck allow
homogeneous and heterogeneous microplate assays to be performed in this
flexible system. The FI and Abs microplate reader with built-in cuvette port
allows easy conversion and optimization of very-low-throughput assays into
medium-high-throughput microplate assays.
The SpectraMax M2 allows faster, more precise results and reagent savings. In addition, combined absorbance and fluorescence intensity assays can be run by issuing a single read command. Within endpoint and kinetic measurement protocols, up to four wavelength pairs can be read at a time, allowing for improved efficiency. Key Features: dual mode cuvette port, robotics compatible, multilabel, dual monochromators, and PathCheck.
Applications include: DNA, RNA, and protein quantitation and purity; PicoGreen, NanoOrange, and Bradford; ELISAs and enzyme kinetics; drug dissolution profiles; live or dead viability or cytotoxicity assays; Caspase-3 and protease assays; cAMP assays using CatchPoint Assay Kits; and low volume applications.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Molecular Devices SpectraMax M2 Multilabel Microplate Reader includes:
- > Windows 10 Data Station Pre-Loaded with SoftMax Pro Software 6.3
- > Operator's Guide
- > Interface Cable
- > Molecular Devices Absorbance and Fluorescence Validation Report
- > Top Reading Plate
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Molecular Devices SpectraMax M2 Multilabel Microplate Reader Specifications
Absorbance Photometric Performance
|
Wavelength Selection
|
Monochromator, tunable in 1 nm increments
|
Wavelength Range
|
200 to 1000 nm
|
Wavelength Bandwidth
|
≤ 4.0 nm
|
Wavelength Accuracy
|
± 2.0 nm
|
Wavelength Repeatability
|
± 0.2 nm
|
Photometric Range
|
0 to 4.0 OD
|
Photometric Resolution
|
0.001 OD
|
Photometric Accuracy (Microplate)
|
< ± 0.006 OD ± 1%, 0 to 2 OD
|
Photometric Accuracy (Cuvette)
|
< ± 0.005 OD ± 1%, 0 to 2 OD
|
Photometric Precision
|
< ± 0.003 OD ± 1%, 0 to 2 OD
|
Baseline Flatness
|
< 0.001 OD
|
Stray Light
|
< 0.05% @ 230 nm
|
Fluorescence Photometric Performance
|
Dual Monochromators
|
1 nm increment selection
|
EX Wavelength Range
|
250 to 850 nm
|
EM Wavelength Range
|
250 to 850 nm
|
Wavelength Bandwidth (EX, EM)
|
9 nm
|
Top-Read Optimized Sensitivity
|
5 pM (fluorescein) in 96 wells
|
Top-Read Guaranteed Sensitivity
|
15 pM (fluorescein) in 96 wells
|
Time-Resolved Fluorescence (secondary mode)
|
Wavelength Range
|
360 to 850 nm
|
Data Collection
|
50 to 1450 usec., 200 usec. increments
|
Guaranteed Sensitivity
|
7 pM Eu-chelate
|
Luminescence (secondary mode)
|
|
Wavelength Range
|
360 to 850 nm
|
Guaranteed Sensitivity
|
50 fM alkaline phosphatase
|
General Photometric Performance
|
Plate Formats
|
6, 12, 24, 48, 96, 384 wells
|
Light Source
|
Xenon flash lamp (1 joule/flash)
|
Detector
|
Photomultiplier tube (PMT)
|
Read Time
|
96 wells: Abs 18 sec., Fl 15 sec.
384 wells: Abs 49 sec., Fl 45 sec.
|
Shaker Time
|
0 to 999 seconds
|
Temperature Control
|
Ambient +4C to 45C
|
Temperature Uniformity
|
< 1C at 37C set point
|
Temperature Accuracy
|
± 1C at 37C set point
|
General Specifications
|
Dimensions (H x W x D)
|
22 x 58 x 38 cm (8.6 x 22.8 x 15 in)
|
Weight
|
15.9 kg (25 lbs)
|
Computer Specifications
|
Model
|
Hewlett-Packard
|
Processor
|
Intel Core I5-7200U
CPU
|
RAM
|
8GB
|
Hard Drive
|
240GB Internal SSD
|
Operating System
|
Windows 10 Professional, 64-bit
|
- Confirm the power supply and PCB output correct, stable voltages across all required connections
- Inspect the display for any dim or dead pixels; replace affected components to maintain full visual clarity and operational accuracy
- Verify full functionality of all buttons, ensuring precise and responsive input control
- Check the XY assembly for proper movement and alignment; thoroughly clean all railings, removing any debris or contaminants
- Confirm stable software communication and connection between the instrument and PC, addressing any connectivity issues
- Inspect and clean or replace all filters as required to ensure optimal airflow and performance
- Examine internal wiring and electrical connections for any breaks, disconnections, or tears; replace or secure any compromised wiring
- Test the lamp’s full functionality, replacing it if any performance issues, dimming, or faults are detected
- Execute the necessary validation protocol, documenting completion with a date on a calibration/verification certificate
- Perform and verify appropriate updates to the PC operating system for compatibility and security
- Conduct a thorough cosmetic inspection, ensuring all external components meet quality and aesthetic standards
Technician Inspection and Recertification Checklist: Instruments
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Confirm stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim or dead pixels; replace as needed
|
Completed
|
|
Verify functionality of all buttons for precise, responsive control
|
Completed
|
|
Check XY assembly for proper movement; clean railings to remove debris
|
Completed
|
|
Confirm stable software communication between instrument and PC
|
Completed
|
|
Inspect and clean or replace filters to maintain optimal performance
|
Completed
|
|
Examine internal wiring for damage; replace or secure compromised connections
|
Completed
|
|
Test lamp functionality; replace if dimming or faults are detected
|
Completed
|
|
Execute validation protocol and document on calibration/verification certificate
|
Completed
|
|
Update PC operating system for compatibility and security
|
Completed
|
|
Conduct cosmetic inspection to ensure quality of external components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|