|
Mettler Toledo PM4600 Balance
The Mettler Toledo PM-4600 Balance is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
The Mettler Toledo PM4600 Balance includes:
- > Power Cord
- > Calibration Label
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Mettler Toledo PM4600 Balance Specifications
Capacity
|
4100 Grams
|
Weighing ranges
|
600.00g/4100.00g
|
Corresponding Readability
|
0.01g/0.1g
|
Reproducibility
|
0.01g/0.03g
|
Tare
|
Full Range
|
Stabilization time (selectable)
|
1.5/2/3s
|
Selectable Weight Units
|
Grams, kilograms, carats, pounds, ounces, etc.
|
Features
|
Bidirectional data interface
|
Mettler Toledo PM4600 Manual Click Here to View all the Technical Specs for the PM 4600
- Verify power supply output for precise voltage delivery in accordance with equipment specifications
- Assess full functionality of display and control buttons, replacing any defective or deteriorating components as necessary
- Confirm balance is properly leveled to ensure measurement accuracy and stability
- Examine windscreen glass for cracks or chips, replacing if structural integrity is compromised
- Inspect all internal wiring and electrical connections for secure attachment and check for any loose connections or damaged connectors
- Check power cord integrity, ensuring there are no frays, cuts, or damage affecting function or safety
- Execute internal or external calibration to confirm accuracy of weighing measurements
- Review cosmetic condition, confirming no visible damage or wear that affects functionality
- Verify successful taring of the weighing pan to zero out prior measurements accurately
- Ensure RS232/printer communication functions correctly, validating connection and data transmission (if applicable)
Technician Inspection and Recertification Checklist: Balances
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Verify power supply output voltage
|
Completed
|
|
Assess display and control button functionality, replacing faulty components
|
Completed
|
|
Confirm balance leveling for measurement accuracy
|
Completed
|
|
Inspect windscreen glass for cracks, replace if necessary
|
Completed
|
|
Check internal wiring and electrical connections for security
|
Completed
|
|
Inspect power cord for frays or damage
|
Completed
|
|
Execute internal/external calibration for measurement accuracy
|
Completed
|
|
Review cosmetic condition for visible damage
|
Completed
|
|
Verify weighing pan tares successfully
|
Completed
|
|
Ensure RS232/printer communication is functional (if applicable)
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|