Mettler Toledo MS403S Precision Balance
With a weighing capacity of up to 420 grams and the versatility to weigh objects as light as 0.2 grams with 0.001 gram readability, the MS403S is a great balance for high precision and versatile weighing. Equipped with MonoBloc weighing technology, Mettler Toledos high performance weighing cell, the balance offers a winning combination of fast settling times (1.5 second stabilization time), and consistently accurate and reliable results. The balance also comes equipped with fully automatic time and temperature controlled internal adjustment (FACT) technology, guaranteeing fast, accurate, and precise results in any weighing environment (3 ppm/C sensitivity drift). The housing is constructed out of high grade die cast aluminum which protects the balance and its sensitive sensors from any chemicals, environmental influences, and impacts. Installation is quick and easy with the adjustable leveling feet and the clearly visible level indicator placed in the front of the balance. Operation of the balance is made user friendly by a high contrast display with big numbers, clear symbols and an intuitive menu enabling users to operate the balance quickly and easily. And with 3 freely programmable SmartKeys, routine tasks can be started at the touch of a button. If you require data transfer of your results, the balance is equipped with RS232 and USB ports, easily connecting you to a pc, printer, or bar code reader. When finished with operation of the balance, clean up is made quick and easy through the QuickLock draft shield. The Mettler Toledo MS403S Precision Balance is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
The Mettler Toledo MS403S Precision Balance includes:
- > Power Cord
- > Calibration Report
- > Full 180 Day Parts and Labor Warranty
Mettler Toledo MS403S Precision Balance Specifications
Max. Capacity
|
420 g
|
Readability
|
0.001 g
|
Repeatability
(sd)
|
0.001 g
|
Linearity
|
0.002 g
|
Calibration
Type
|
Internal
|
RS-232
|
Yes
|
Stabilization
Time
|
1.5 seconds
|
Electrical
Requirements
|
100 - 240
VAC, 50/60 Hz
|
Weighing
Units
|
G, kb, mg,
ct,lb, oz, ozt, GN, dwt, mom, msg, tlh, tls, tlt, tola, baht
|
Languages
|
English,
German, French, Spanish, Italian, Polish, Czech, Hungarian
|
- Verify power supply output for precise voltage delivery in accordance with equipment specifications
- Assess full functionality of display and control buttons, replacing any defective or deteriorating components as necessary
- Confirm balance is properly leveled to ensure measurement accuracy and stability
- Examine windscreen glass for cracks or chips, replacing if structural integrity is compromised
- Inspect all internal wiring and electrical connections for secure attachment and check for any loose connections or damaged connectors
- Check power cord integrity, ensuring there are no frays, cuts, or damage affecting function or safety
- Execute internal or external calibration to confirm accuracy of weighing measurements
- Review cosmetic condition, confirming no visible damage or wear that affects functionality
- Verify successful taring of the weighing pan to zero out prior measurements accurately
- Ensure RS232/printer communication functions correctly, validating connection and data transmission (if applicable)
Technician Inspection and Recertification Checklist: Balances
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Verify power supply output voltage
|
Completed
|
|
Assess display and control button functionality, replacing faulty components
|
Completed
|
|
Confirm balance leveling for measurement accuracy
|
Completed
|
|
Inspect windscreen glass for cracks, replace if necessary
|
Completed
|
|
Check internal wiring and electrical connections for security
|
Completed
|
|
Inspect power cord for frays or damage
|
Completed
|
|
Execute internal/external calibration for measurement accuracy
|
Completed
|
|
Review cosmetic condition for visible damage
|
Completed
|
|
Verify weighing pan tares successfully
|
Completed
|
|
Ensure RS232/printer communication is functional (if applicable)
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|