Mass Range
10 – 1350 m/z
Mass Accuracy
± 0.13 u within the calibrated mass range in scan mode
Mass Axis Stability
The mass drift does not exceed the larger absolute value of ± 0.1 Da or ± 100 ppm of measured mass over a 12 hour period at constant temperature ± 3 degrees centrigrade.
Scan Speed
10,000 u/s in fast mode and 2500 u/s in normal mode
SIM Sensitivity
ESI at 400 μ/min or APCI at 1000 μ/min Selected ion monitoring of m/z 609.3 Positive ionization 1 pg reserpine, 100:1 RMS (20:1 peak-to-peak)
Scan Sensitivity
ESI at 400 μ/min or APCI at 1000 μ/min Scan range m/z 100 – 650 Scan speed 2500 u/s Extracted ion at m/z 609.3 Positive ionization 50 pg reserpine, 100:1 RMS (20:1 peak-to-peak)
Flow Rate
ESI: Up to 1.0 ml/min water without a flow splitter APCI: Up to 1.5 ml/min water without a flow splitter
Multisignal acquisition
Ability to cycle through four acquisition modes on a scan-by-scan basis within a single run. Up to four analysis modes can be viewed in the software window and can be linked together to facilitate processing and analysis. Modes of operation: positive/negative polarity switching, variable fragmentor voltage selection, and concurrent SIM/scan.
Operating Conditions
Temperature range: 15 °C to 35 °C (59 °F to 95 °F) Relative humidity: less than 95 % (non-condensing)
Dimensions
Height: 65 cm (26 in) Width: 39 cm (15 in) Depth: 75 cm (29 in) Weight: 50 kg (110 pounds)
Electrical Requirements
200-240V, 50/60Hz
m/z 10 - 1350
ESI at 400 u/min or APCI at 1000 u/min Selected ion monitoring of m/z 609.3 Positive ionization 1 pg reserpine, 100:1 RMS (20:1 peak-to-peak)
ESI at 400 u/min or APCI at 1000 u/min Scan range m/z 100 - 650 Scan speed 2500 u/s Extracted ion at m/z 609.3 Positive ionization 50 pg reserpine, 100:1 RMS (20:1 peak-to-peak)