MJ Research PTC-240 Gradient Thermal Cycler
The MJ Research PTC-240 Gradient Thermal Cycler has a
thermal accuracy of
0.3C of programmed target at 90C and a thermal
uniformity of
0.4C well-to-well within 30 seconds of arrival at 90C. With a
expeditious ramping speed of up to 3C per second and an equipped sample capacity of four
96 well alpha blocks. The PTC-240 can accommodate multiple blocks, such as 384 well, 60 well, 96 well, and dual 48 well blocks, among many others, including specialty blocks like the insitu. The MJ Research PTC-240 Gradient Thermal Cycler is ideal
for labs with high throughput PCR applications. Ports include one 25-pin 8-bit
parallel interface printer port and one RS-232 serial port.
The MJ Research PTC-240 is one of
the most desired thermal cycling systems on the market to date. This robust PCR
has the ability to run four blocks simultaneously. The MJ Research PTC-240
Gradient Thermal Cycler is also available with different alpha block options.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The MJ Research PTC-240 Gradient Thermal Cycler includes:
- > Four 96 Well Alpha Blocks
- > Power Supply
- > Calibration Report
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
MJ Research PTC-240 Gradient Thermal Cycler Specifications
Thermal Accuracy
|
0.3C of programmed target at 90C
|
Thermal Uniformity
|
0.4C well-to-well within 30 seconds of arrival at 90C
|
Ramping Speed
|
Up to 3C/sec
|
Sample Capacity
|
Four Alpha Block, Holds 96 x 0.2 ml tubes or one 96-well
microplate/Block
|
Line Voltage
|
10A @ 200-240 VAC, 50-60Hz
|
Ports
|
One 25-pin 8-bit parallel interface printer port & RS-232 serial
port
|
MJ Research PTC-240 Manual
We also carry the PTC-100, PTC-150, PTC-200, and PTC-225.
Click Here to View all the Technical Specs for the PTC-240
- Validate that the power supply and PCB consistently deliver accurate, stable voltages across all circuits and connections, confirming operational reliability
- Thoroughly inspect the display for any dim, flickering, or dead pixels, replacing compromised segments to ensure clear visibility and precise data display
- Verify complete functionality and responsiveness of all buttons and knobs, confirming accurate and reliable input across all controls
- Ensure the machine completes all internal diagnostic checks with zero errors, indicating system readiness and functional integrity
- Confirm the machine accurately identifies attached blocks (96-well, dual 48, etc.), with seamless recognition across different configurations
- Conduct a detailed inspection of internal wiring and all electrical connections, checking for any signs of breaks, disconnections, or tears; replace or repair any underperforming components
- Examine the connector between the block and base for bent or broken pins, replacing as needed to maintain secure, reliable connections
- Calibrate well temperatures with a NIST-traceable thermometer, verifying accuracy and uniformity across all wells to meet specification standards
- Perform a comprehensive cosmetic inspection, ensuring all exterior surfaces and components are free of damage, wear, or defects, maintaining aesthetic and professional quality
Technician Inspection and Recertification Checklist: Thermal Cycler
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, accurate voltage outputs from power supply and PCB across circuits
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed for clarity
|
Completed
|
|
Verify functionality and responsiveness of all buttons and knobs
|
Completed
|
|
Confirm machine completes internal diagnostics with zero errors
|
Completed
|
|
Ensure accurate block identification across configurations (96-well, dual 48, etc.)
|
Completed
|
|
Inspect internal wiring and connections for integrity; repair or replace as needed
|
Completed
|
|
Check connector between block and base for bent or broken pins; replace as necessary
|
Completed
|
|
Calibrate well temperatures with a NIST-traceable thermometer for accuracy
|
Completed
|
|
Conduct cosmetic inspection to ensure quality and professionalism of all surfaces
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|