MJ Research PTC-200 Thermal Cycler Dual 48
Thermal Range 5C to 105C - no more than 30C below ambient
Sample Capacity Dual 48 Well Alpha Block
- Dual 48-well block design allows two independent programs to run simultaneously
- Hot Bonnet heated lid controls pressure and height for oil-free cycling
- Exceptional thermal accuracy of ±0.3C at 90C ensures reliable performance
- Unparalleled temperature uniformity of ±0.4C within 30 seconds for consistent results
- Intuitive software with an easy-to-read 20 x 4 LCD display for seamless programming and monitoring
- Supports calculated sample temperature control, block control, or probe control for protocol flexibility
- Instant Incubate feature enables continuous-temperature incubations for extended applications
- Compact, space-saving design makes setup and transportation easy
The MJ Research PTC-200 Thermal Cycler Dual 48 utilizes interchangeable sample blocks and heat pump assemblies allow the user to change format easily between runs while retaining NIST-traceable accuracy. The PTC-200 has a dual 48-well alpha block, allowing two independent programs to be run at the same time, and can accommodate 60 well blocks, 96 well blocks, 384-well blocks, and many others, including specialty blocks like the insitu. The precise Peltier effect heat pumps deliver the full temperature range of 0 to 100C electronically, all in a compact, small-footprint instrument. The MJ Research PTC-200 offers speedy ramping rates of up to 3.0/sec in models that hold 0.2ml tubes or V-well plates. It also features unparalleled temperature homogeneity of 0.4C within 30 seconds of arrival at 60C for all standard models. The PTC-200 has a large memory and amiable software that allows control over virtually every thermal parameter.
The MJ Research PTC-200 is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
The MJ Research PTC-200 Thermal Cycler Dual 48 includes:
- > Dual 48 Well Alpha Block
- > NIST Traceable Calibration Certificate
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
MJ Research PTC-200 Thermal Cycler Dual 48 Specifications
Thermal Range
|
5C to 105C, but no more than 30C below ambient temperature (4 to 100C for the slide Chambers Alpha Unit)
|
Thermal Accuracy
|
± 0.3C of programmed target at 90C- NIST-traceable
|
Thermal Uniformity
|
± 0.4C well-to-well within 30 seconds of arrival at 90C (for most Alpha units; see specifications for individual Alpha Units)
|
Ramping Rate
|
Up to 3˚C/sec for all single- and dual-block Alpha units; 1.2˚C/sec for the Twin Tower
|
Gradient Function
|
No
|
Memory
|
200 typical programs in up to 12 individual folders
|
Sample Capacity
|
Dual 48 Well Alpha Block
|
Parrallel
|
1x 25-pin 8-bit parallel interface printer port
|
Serial
|
1x 9-pin RS-232 serial port for printer or remote use
|
IEEE
|
one IEEE-488 bidirectional general purpose interface bus
|
Display
|
one 20 x 4 LCD alphanumeric display
|
Dimensions (L x W x H)
|
12 x 12 x 12 in. (30.5 x 30.5 x 30.5 cm)
|
Weight
|
25 lb. (11.3 kg)
|
Electrical Requirements
|
100 - 240 VAC, 50-60Hz, 850W
|
- Validate that the power supply and PCB consistently deliver accurate, stable voltages across all circuits and connections, confirming operational reliability
- Thoroughly inspect the display for any dim, flickering, or dead pixels, replacing compromised segments to ensure clear visibility and precise data display
- Verify complete functionality and responsiveness of all buttons and knobs, confirming accurate and reliable input across all controls
- Ensure the machine completes all internal diagnostic checks with zero errors, indicating system readiness and functional integrity
- Confirm the machine accurately identifies attached blocks (96-well, dual 48, etc.), with seamless recognition across different configurations
- Conduct a detailed inspection of internal wiring and all electrical connections, checking for any signs of breaks, disconnections, or tears; replace or repair any underperforming components
- Examine the connector between the block and base for bent or broken pins, replacing as needed to maintain secure, reliable connections
- Calibrate well temperatures with a NIST-traceable thermometer, verifying accuracy and uniformity across all wells to meet specification standards
- Perform a comprehensive cosmetic inspection, ensuring all exterior surfaces and components are free of damage, wear, or defects, maintaining aesthetic and professional quality
Technician Inspection and Recertification Checklist: Thermal Cycler
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, accurate voltage outputs from power supply and PCB across circuits
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed for clarity
|
Completed
|
|
Verify functionality and responsiveness of all buttons and knobs
|
Completed
|
|
Confirm machine completes internal diagnostics with zero errors
|
Completed
|
|
Ensure accurate block identification across configurations (96-well, dual 48, etc.)
|
Completed
|
|
Inspect internal wiring and connections for integrity; repair or replace as needed
|
Completed
|
|
Check connector between block and base for bent or broken pins; replace as necessary
|
Completed
|
|
Calibrate well temperatures with a NIST-traceable thermometer for accuracy
|
Completed
|
|
Conduct cosmetic inspection to ensure quality and professionalism of all surfaces
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
 |