MJ Research PTC-150 Thermal Cycler
The MJ Research PTC-150 Mini Thermal Cycler is an easy-to-use, compact, reliable thermal scanner with Hot Bonnet heated lid for oil-free cycling. The PTC-150 features Instant Incubate, which allows for continuous hassle-free temperature incubations. All of our equipment is fully tested to perform at factory specifications before being shipped.
The MJ Research PTC-150 Thermal Cycler includes:
- > Power Cord
- > NIST Traceable Calibration Certificate
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
MJ Research PTC-150 Thermal Cycler Specifications
Thermal Range
|
-9 to 105C (no more than 35 below ambient temperature)
|
Accuracy
|
0.3C of programmed target at 90C, NIST traceable
|
Thermal Homogeneity
|
0.4C well-to-well within 12 seconds of arrival at 90C
0.3C well-to-well within 40 seconds of arrival at 90C
|
Sample Capacity
|
16-well block: 16 x 0.5mL tubes
25-well block: 25 x 0.2mL tubes or one 25-well microplate
|
Ports
|
One 25-pin 8-bit parallel interface printer port
|
Memory
|
80 typical programs, in nonvolatile memory each program may have up
to 100 steps; maximum of 400 steps among all programs
|
MJ Research PTC-150 Thermal Cycler ManualClick Here to View all the Technical Specs for the PTC-150
- Validate that the power supply and PCB consistently deliver accurate, stable voltages across all circuits and connections, confirming operational reliability
- Thoroughly inspect the display for any dim, flickering, or dead pixels, replacing compromised segments to ensure clear visibility and precise data display
- Verify complete functionality and responsiveness of all buttons and knobs, confirming accurate and reliable input across all controls
- Ensure the machine completes all internal diagnostic checks with zero errors, indicating system readiness and functional integrity
- Confirm the machine accurately identifies attached blocks (96-well, dual 48, etc.), with seamless recognition across different configurations
- Conduct a detailed inspection of internal wiring and all electrical connections, checking for any signs of breaks, disconnections, or tears; replace or repair any underperforming components
- Examine the connector between the block and base for bent or broken pins, replacing as needed to maintain secure, reliable connections
- Calibrate well temperatures with a NIST-traceable thermometer, verifying accuracy and uniformity across all wells to meet specification standards
- Perform a comprehensive cosmetic inspection, ensuring all exterior surfaces and components are free of damage, wear, or defects, maintaining aesthetic and professional quality
Technician Inspection and Recertification Checklist: Thermal Cycler
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, accurate voltage outputs from power supply and PCB across circuits
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed for clarity
|
Completed
|
|
Verify functionality and responsiveness of all buttons and knobs
|
Completed
|
|
Confirm machine completes internal diagnostics with zero errors
|
Completed
|
|
Ensure accurate block identification across configurations (96-well, dual 48, etc.)
|
Completed
|
|
Inspect internal wiring and connections for integrity; repair or replace as needed
|
Completed
|
|
Check connector between block and base for bent or broken pins; replace as necessary
|
Completed
|
|
Calibrate well temperatures with a NIST-traceable thermometer for accuracy
|
Completed
|
|
Conduct cosmetic inspection to ensure quality and professionalism of all surfaces
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|