Labnet VorTemp 56 Incubator Shaker with Microplate Platform
Max Speed 15,200 RPM (25,830 x g RCF)
Temperature Range -10C to 40C
- Combines incubation and shaking functions in a single compact unit
- Microprocessor-controlled interface for precise adjustments and digital display
- 3mm orbit ensures thorough sample mixing and aeration
- Selectable acceleration rates provide controlled start-up
- Timer function allows operation up to 99 minutes or continuous mode
- Heavy-gauge, enamel-coated steel construction for durability and corrosion resistance
- Stable base design ensures vibration-free operation, even at high speeds
- Straightforward controls simplify parameter adjustments for efficient operation
The Labnet VorTemp 56 Incubator Shaker with Microplate Platform is a compact benchtop unit designed to combine precise incubation and vigorous shaking in one device. Ideal for biological, biochemical, and clinical laboratories, it efficiently handles microtubes and microplates while conserving bench space. The microprocessor-controlled interface allows for precise adjustments of temperature, speed, and time, all displayed digitally for ease of use. Shaking speeds range from 200 to 1,200 RPM with a 3mm orbit, ensuring thorough mixing and aeration. A built-in timer function allows for operation up to 99 minutes or continuous use for extended protocols. Constructed from heavy-gauge, enamel-coated steel, the VorTemp 56 provides a stable, vibration-free operation, even at high speeds. With its intuitive controls and reliable performance, this incubator shaker is a versatile solution for labs performing routine thermal shaking applications.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Labnet VorTemp 56 Incubator Shaker with Microplate Platform includes:
- > Incubator Shaker
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Labnet VorTemp 56 Incubator Shaker Specifications
Temperature Range
|
Ambient +5 to 99 C
|
Temperature Accuracy
|
0.5 C
|
Speed Range
|
200 to 1,200 RPM
|
Motion/Orbit
|
Circular/3mm
|
Timer
|
1 to 99 minutes or continuous
|
Capacity
|
56 x 1.5/2.0 mL tubes; 4 standard or 2 deep well microplates
|
Dimensions (W x D x H)
|
9 x 12.3 x 9 in (23 x 31 x 23 cm)
|
Electrical Requirements
|
115VAC, 50/60Hz, 4.8A
|
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Examine chamber glass for cracks or chips, replace as needed | Completed | |
Inspect door seals for wear, replace to ensure tight seal | Completed | |
Check display brightness, replace dim or dead pixels | Completed | |
Test all buttons for full functionality and responsiveness | Completed | |
Replace platform assembly bearings for reliable motion | Completed | |
Inspect drive belt for wear, replace if torn or cracked | Completed | |
Examine wiring for frays or breakages, repair as needed | Completed | |
Verify PCB test points, replace faulty components | Completed | |
Check drive belt tension for proper alignment | Completed | |
Confirm compressor functionality, clean filter and fins | Completed | |
Calibrate internal temperature with NIST probe | Completed | |
Calibrate platform RPM with NIST-traceable tachometer | Completed | |
Conduct cosmetic inspection for appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|