Labnet VorTemp 56 Incubator Shaker
Temperature Range Ambient +5 to 99 C
Speed Range 200 to 1,200 RPM
Motion / Orbit
Circular / 3mm
- Adjustable temperature range suitable for various applications, including bacterial cultures in tubes and microplates.
- Mechanical convection ensures a uniform, stable environment in the chamber.
- Usable in cold rooms for sub-ambient applications.
- Simultaneous shaking and incubation of samples with intelligent control for precise operation.
- Adjustable shaking speed and acceleration rates to protect delicate samples.
- Microprocessor-controlled parameters, including temperature and timed operation.
- Self-diagnostic software for monitoring temperature and shaking speed, with error alerts.
- Feedback loops for accurate maintenance of set parameters, regardless of load or voltage changes.
- Sturdy construction provides stability during high-speed shaking.
- Insulated chamber for temperature maintenance and reduced noise.
The Labnet VorTemp 56 Incubator Shaker combines high-speed shaking with precise temperature control, making it ideal for bacterial cultures, biochemical applications, and routine lab procedures. With a temperature range from 5°C above ambient to 99°C and an impressive 0.5°C accuracy, this shaker provides dependable environmental control for samples in tubes and microplates. It offers adjustable speeds from 200 to 1,200 RPM on a 3mm orbit, with three acceleration rates to protect delicate samples.
Designed for robust performance, the VorTemp 56 features mechanical convection for a uniform chamber environment, microprocessor-controlled settings, and self-diagnostic software that alerts users to temperature or speed discrepancies. The unit accommodates up to 56 x 1.5/2.0 mL tubes or 4 standard microplates, making it versatile for a variety of laboratory tasks. Sturdy construction, an insulated chamber, and feedback loops for stability during high-speed shaking round out its intelligent design, delivering reliable results across a range of applications.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Labnet VorTemp 56 Incubator Shaker includes:
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Labnet VorTemp 56 Incubator Shaker Specifications
Temperature Range
|
Ambient +5 to 99 C
|
Temperature Accuracy
|
0.5 C
|
Speed Range
|
200 to 1,200 RPM
|
Motion/Orbit
|
Circular/3mm
|
Timer
|
1 to 99 minutes or continuous
|
Capacity
|
56 x 1.5/2.0 mL tubes; 4 standard or 2 deep well microplates
|
Dimensions (W x D x H)
|
9 x 12.3 x 9 in (23 x 31 x 23 cm)
|
Electrical Requirements
|
115VAC, 50/60Hz, 4.8A
|
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Examine chamber glass for cracks or chips, replace as needed | Completed | |
Inspect door seals for wear, replace to ensure tight seal | Completed | |
Check display brightness, replace dim or dead pixels | Completed | |
Test all buttons for full functionality and responsiveness | Completed | |
Replace platform assembly bearings for reliable motion | Completed | |
Inspect drive belt for wear, replace if torn or cracked | Completed | |
Examine wiring for frays or breakages, repair as needed | Completed | |
Verify PCB test points, replace faulty components | Completed | |
Check drive belt tension for proper alignment | Completed | |
Confirm compressor functionality, clean filter and fins | Completed | |
Calibrate internal temperature with NIST probe | Completed | |
Calibrate platform RPM with NIST-traceable tachometer | Completed | |
Conduct cosmetic inspection for appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|