Labline 4625 Titer Shaker
Speed Range 40 to 1,100 RPM
Platform Size 10.5 x 7.5 in. (26.7 x 19.1 cm)
- Designed for virology, immunology, serology, and hematology applications
- Can securely hold one to four 96-well titer plates with two plate retaining springs
- Features a white, non-slip rubber platform mat to enhance sample visibility
- Provides a vortexing motion for thorough and even mixing of every well
- Offers a solid-state speed control for shaking speeds from 40 to 1100 rpm with a 3mm orbit
- Includes a built-in electrical timer for continuous operation or timed cycles up to 5 minutes
- Rugged powder-coated steel body with four rubber feet for stability
- Equipped with a 10.5 x 7.5 inch (26.7 x 19.1 cm) platform
The Labline 4625 Titer Shaker is reliable and dependable, combining top-notch and uncompromising quality with an affordable price. The Barnstead Titer Plate Shaker is ideal for virology, immunology, serology and hematology applications. Two plate retaining springs on platform firmly hold one to four 96-well titer plates. White, non-slip rubber platform mat increases sample visibility. Vortexing motion thoroughly mixes every well evenly. Solid-state speed control provides accurate, continuous regulation of shaking speeds from 40 to 1100 rpm with a 3mm orbit. Built-in electrical timer can be set for continuous operation or timed cycles of up to 5 minutes. Unit shuts off automatically at end of time cycle. Four finger slots provide for easy plate removal. Rugged powder-coated steel body sits on four rubber feet. (Barnstead 4625). Equipped with a 10.5x7.5 inch platform.
All of our equipment will be fully tested to perform at factory specifications before being shipped.
The Labline 4625 Titer Shaker includes:
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Labline 4625 Titer Shaker Specifications
Speed Range
|
40 to 1,100 RPM
|
Shaker Motion
|
0.12 in orbit
|
Timer
|
Up to 5 min in 1 min increments or continuous
|
Overall Dimensions (W x D x H)
|
11 x 11 x 7 in (28 x 28 x 18 cm)
|
Electrical Requirements
|
120 VAC, 60 Hz, 120 W, 1 A
|
Labline 4625 Titer Shaker Manual
Click Here to View all the Technical Specs for the 4625 Titer Shaker
Labline 4625 Titer Shaker
- Validate that the power supply and PCB deliver stable, correct voltage outputs across all necessary connections, ensuring reliable operation
- Thoroughly inspect the display for any dim, flickering, or dead pixels; replace affected elements to maintain full readability and operational clarity
- Test the full range of button functionality, confirming responsive, accurate input recognition across all settings
- Conduct a detailed inspection of bearings and bearing housing, identifying any worn or damaged bearings, and replace as needed to prevent operational disruption
- Examine the drive belt for signs of cracking, fraying, or material breakdown; replace as required to ensure sustained performance
- Measure and adjust motor drive belt tension to optimal specifications, securing reliable and efficient power transmission
- Verify accuracy of the onboard timer, confirming proper functionality and replacing if it deviates from expected performance
- Calibrate shaker rotation speed with a NIST-traceable tachometer, validating precision within specified operational limits
- Run the shaker continuously for an 8-12 hour testing period, closely observing the shaking platform for smooth, consistent motion, while monitoring for any error indicators such as knocking, skipping, or unusual vibrations
- Perform a comprehensive cosmetic inspection to confirm the device’s exterior meets aesthetic and quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Shakers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed
|
Completed
|
|
Test button functionality for responsiveness and accuracy across all settings
|
Completed
|
|
Inspect bearings and housing for wear; replace as needed to prevent disruption
|
Completed
|
|
Examine drive belt for cracking, fraying, or material breakdown; replace as required
|
Completed
|
|
Measure and adjust motor drive belt tension to optimal specifications
|
Completed
|
|
Verify onboard timer accuracy; replace if it deviates from expected performance
|
Completed
|
|
Calibrate shaker rotation speed using a NIST-traceable tachometer
|
Completed
|
|
Run 8-12 hour continuous test; observe platform for smooth, consistent motion
|
Completed
|
|
Perform comprehensive cosmetic inspection for quality standards compliance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|