Labconco FreeZone 4.5L -50C Benchtop Freeze Dryer System
Ice Holding Capacity 4.5L
Manifold 12 Port Drying Chamber
- Designed for light sample loads, this compact Labconco FreeZone 4.5L Benchtop Freeze Dryer optimizes limited lab space
- Stainless steel collector coil holds 4.5 liters of ice and cools to -50°C, suitable for samples with low eutectic points
- Features dual 1/3 hp HCFC/CFC-free refrigeration systems for eco-friendly and efficient cooling
- Includes a vacuum break valve to protect the vacuum pump by bleeding air into the system during power interruptions
- Moisture sensor prevents refrigeration or vacuum start-up when moisture is detected, protecting system components
- Automatic vacuum pump shutdown activates during leaks to maintain system integrity and prevent damage
- Comes with an Edwards RV5 vacuum pump, 12 port manifold, and essential accessories for complete setup
- Capable of removing up to 2 liters of water over 24 hours with a 12-port drying chamber for efficient lyophilization
- Complies with UL, CAN/CSA, and CE standards for high safety and operational standards
- Includes a 180-day full parts and labor warranty for reliable support and service
The Labconco FreeZone 4.5L -50C Benchtop Freeze Dryer System is designed for lyophilizing light sample loads. The compact design is ideal for labs with limited floor space. The upright stainless steel collector coil holds 4.5 liters of ice before defrosting. Two 1/3 hp HCFC/CFC-free refrigeration systems cool the collector to -50°C, suitable for samples with low eutectic points including acetonitrile.
FreeZone exclusive features protect your vacuum pump from costly damage: A vacuum break valve bleeds air into the system when power to the freeze dryer or vacuum pump is shut off for more than 5 minutes. A moisture sensor prevents refrigeration or vacuum start-up when moisture is detected in the drain line. Vacuum pump is automatically shut down when a leak is present in initial pull down or during run.
The Labconco FreeZone 4.5L -50C Benchtop Freeze Dryer System includes:
- > Edwards RV5 Vacuum Pump
- > 12 Port Manifold
- > Vacuum Tubing and Hardware
- > Mist Filter
- > Operator's Guide
- > Power Cords
- > 180 Day Full Parts & Labor Warranty
Labconco FreeZone 4.5L -50C Benchtop Freeze Dryer System Specifications
Ice Holding Capacity
|
4.5L
|
Lowest Temperature
|
-50C
|
Water Removal Over 24 Hours
|
2L
|
Manifold
|
12 Port Drying Chamber
|
Electrical Requirements
|
115V, 60Hz, 14A, 1Ph
|
Vacuum Pump Displacement Required
|
98L/min
|
Dimensions (W x D x H)
|
16.5 x 22.4 x 18.9 in
|
Weight
|
139 lbs. (63 kg.)
|
Conformance
|
UL, CAN/CSA, CE
|
- Inspect the display for pixel clarity and brightness issues; replace any dim or defective pixels to ensure optimal visual quality.
- Test each button for responsiveness, resistance, and continuity; replace any buttons that show wear or unresponsiveness.
- Check the acrylic/glass lid for structural integrity, looking for cracks or chips; replace if any damage is found.
- Examine all external gasket seals for elasticity, cracks, or tears; replace compromised seals with OEM-approved parts.
- Remove and replace all manifold ports with new OEM parts, ensuring proper fit and sealing to maintain system integrity.
- Drain and replace the oil, verifying the new oil’s viscosity and particulate levels fall within operational standards.
- Conduct diagnostic checks on all PCB test points, replacing any malfunctioning components to maintain circuit integrity.
- Inspect all internal rubber connections and gaskets for cracks or brittleness; replace compromised parts as needed.
- Disassemble and clean the condenser coils and fan assembly to remove dust; replace worn components to ensure efficiency.
- Confirm the start capacitor for the refrigeration system meets the manufacturer-specified capacitance range; replace if readings deviate.
- Test and calibrate all internal sensors and safety switches, ensuring accurate readings and replacing any faulty components.
- Verify displayed temperatures by cross-referencing with independent calibrated probes; recalibrate the system if discrepancies are detected.
- Check vacuum pressure using an independent gauge, aiming for ~0.40 mbar; address any leaks or inefficiencies affecting vacuum levels.
- Confirm the system achieves and maintains low temperatures (~-50°C or ~-80°C) using certified measurement tools; troubleshoot as necessary.
- Inspect all external surfaces, labels, and interfaces for cosmetic wear; address any scratches, blemishes, or visible signs of damage.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Inspect display for pixel issues and replace if needed | Completed | |
Test button functionality and replace any defective buttons | Completed | |
Examine door seal and lock functionality; replace if compromised | Completed | |
Conduct PCB test point diagnostics and replace as needed | Completed | |
Verify compressor and internal fan functionality | Completed | |
Confirm temperature probe accuracy with NIST-traceable probe | Completed | |
Inspect all wiring for frays, tears, or breakages | Completed | |
Test start capacitor capacitance range; replace if needed | Completed | |
Replace all internal filters and clean condenser fins | Completed | |
Test alarm battery for full functionality | Completed | |
Run 72-hour test with data recorder to confirm stable temperatures | Completed | |
Conduct cosmetic inspection and address any issues | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|