Labconco CentriVap Mobile System w/ Heat Boost, Cat # 7812013
The is Labconco Centrivap Mobile System combines a concentrator and cold trap into one compact unit. The concentrator uses centrifugal force, heat, and vacuum to rapidly evaporate solvents. The addition of centrifugation keeps solvents from foaming or bumping, while also concentrating solutes in the tube bottoms. The concentrator is equipped with a 300 watt heater to speed evaporation by warming the chamber during processing. The microprocessor, which controls the operation of the heater and the motor, provides excellent regulation and reproducibility of protocols. Save up to nine protocols for easy repeatability. As an integral part at the CentriVap Mobile System, the cold trap protects the vacuum pump by trapping moisture, vapors and corrosive fumes as they evaporate from the samples.
This unit is in excellent condition, and all of our equipment is fully tested to perform at factory specifications before being shipped.
The Labconco CentriVap Mobile System w/ Heat Boost includes:
- > Fixed Angle Rotor
- > Edwards RV3 Rotary Vane Vacuum Pump
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Labconco CentriVap Mobile System w/ Heat Boost Specifications:
Max. Speed
|
1,725 RPM
|
Temperature Range
|
Up to 100C w/ Heat Boost
|
Time Range
|
1 to 999 minutes
|
Vacuum Level
|
0.002 mbar
|
Electrical Requirements
|
115V, 60Hz, 10.5A
|
Rotor Capacity
|
132 x 12-13mm Tubes
|
Cold Trap Temperature
|
-50C
|
Chamber Material
|
Epoxy-coated cast aluminum
|
Labconco CentriVap Mobile System w/ Heat Boost Manual
- Inspect the display panel, replacing any dim or dead pixels to ensure clear visual feedback
- Verify full functionality of all buttons, confirming consistent responsiveness
- Examine the acrylic or glass lid for any cracks or chips, ensuring safety and integrity
- Check the vacuum chamber seal for any tears or cracks that could impact pressure retention
- Test chamber lid functionality, confirming all micro switches report a consistent .04 Ohms
- Verify external connections to the pump or cold trap, ensuring they are properly sealed and leak-free
- Inspect and clean the internal fan and condenser (if applicable) for efficient cooling and airflow
- Examine all internal wiring for frays or breakages, confirming secure and intact connections
- Test motor functionality, replacing motor or bearings if performance is compromised
- Inspect internal pump seals, replacing any worn or damaged seals to maintain vacuum integrity
- Inspect all internal hoses for potential leaks, ensuring tight, clean seals
- Confirm unit is capable of achieving all manufacturer-specified temperature ranges
- Verify the unit can reach specified vacuum levels in accordance with manufacturer standards
- Conduct a 6-8 hour cycle test to confirm consistent operational performance
Technician Inspection and Recertification Checklist: Centrifugal Evaporators
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display panel for dim or dead pixels
|
Completed
|
|
Confirm functionality of all buttons
|
Completed
|
|
Check acrylic/glass lid for cracks or chips
|
Completed
|
|
Inspect vacuum chamber seal for tears or cracks
|
Completed
|
|
Confirm chamber lid micro switches report .04 Ohms
|
Completed
|
|
Verify pump or cold trap connections are sealed and leak-free
|
Completed
|
|
Inspect and clean internal fan and condenser (if applicable)
|
Completed
|
|
Inspect internal wiring for frays or breakages
|
Completed
|
|
Test motor functionality, replace motor/bearings if needed
|
Completed
|
|
Inspect internal pump seals, replace if needed
|
Completed
|
|
Inspect all internal hoses for leaks, ensure clean seals
|
Completed
|
|
Confirm unit achieves specified temperature ranges
|
Completed
|
|
Verify unit reaches specified vacuum levels
|
Completed
|
|
Conduct a 6-8 hour cycle test for operational performance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|