Labconco CentriVap Complete Vacuum Concentrator
The Labconco CentriVap Complete Vacuum Concentrator (Catalog #: 7315022) is designed to rapidly evaporate and condense solvents from biological and analytical samples from the use of centrifugal force combined with heat and vacuum. The CentriVap Complete system combines a concentrator, cold trap, and a vacuum pump in one cabinet. Centrifugation removes bumping and foaming as vacuum is applied while also concentrating the solute within the vial, allowing recovery of solutes from volumes as low as a few microliters. The concentrator is also equipped with a 300 watt heater to speed evaporation by warming the chamber during processing. A timed outlet for the vacuum pump connection delays the start of the vacuum pump until the rotor reaches optimal operating speed. As an integral part of the CentriVap Complete system, the cold trap can be used to trap moisture, vapors, and corrosive fumes as they evaporate from the samples.
Equipped with a safety switch the Labconco CentriVap Complete Vacuum Concentrator prevents the lid from being opened while the rotor is in operation. The CentriZap strobe light is also included, which allows you to safely view the samples while the unit is running, without stopping the rotor and lifting the lid to check dryness. The compact, space saving design allows the system to be operable on a benchtop or other laboratory work surface. The microprocessor, which controls the operation of the heater
and the motor, provides excellent regulation and reproducibility of protocols. The memory has capability to store up to nine different protocols so exact test factors can be
easily repeated.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Labconco CentriVap Complete Vacuum Concentrator includes:
- > 7315022 CentriVap Concentrator
- > 7455100 12-13 mm Fixed-Angle Rotor
- > Internal Vacuum Pump
- > Glass Jar
- > CentriZap Strobe Light
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Labconco CentriVap Complete Vacuum Concentrator Specifications
Max Speed
|
1,725 RPM
|
Temperature Range
|
Up to 100C
|
Time Range
|
1 to 999 minutes
|
Vacuum Level
|
7 mbar
|
Rotor Capacity
|
1.5/2.0 ml
|
Heater Power
|
300W
|
Dimensions (W x D x H)
|
26.1 x 22.7 x 16.9 in (66.3 x 57.7 x 42.9 cm)
|
Weight
|
145 lbs (66 kg)
|
Electrical Requirements
|
115V, 60Hz, 12A
|
Labconco CentriVap Complete Vacuum Concentrator Manual
Click Here to View the Technical Specs for the Labconco CentriVap Complete Vacuum Concentrator
- Inspect the display panel, replacing any dim or dead pixels to ensure clear visual feedback
- Verify full functionality of all buttons, confirming consistent responsiveness
- Examine the acrylic or glass lid for any cracks or chips, ensuring safety and integrity
- Check the vacuum chamber seal for any tears or cracks that could impact pressure retention
- Test chamber lid functionality, confirming all micro switches report a consistent .04 Ohms
- Verify external connections to the pump or cold trap, ensuring they are properly sealed and leak-free
- Inspect and clean the internal fan and condenser (if applicable) for efficient cooling and airflow
- Examine all internal wiring for frays or breakages, confirming secure and intact connections
- Test motor functionality, replacing motor or bearings if performance is compromised
- Inspect internal pump seals, replacing any worn or damaged seals to maintain vacuum integrity
- Inspect all internal hoses for potential leaks, ensuring tight, clean seals
- Confirm unit is capable of achieving all manufacturer-specified temperature ranges
- Verify the unit can reach specified vacuum levels in accordance with manufacturer standards
- Conduct a 6-8 hour cycle test to confirm consistent operational performance
Technician Inspection and Recertification Checklist: Centrifugal Evaporators
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display panel for dim or dead pixels
|
Completed
|
|
Confirm functionality of all buttons
|
Completed
|
|
Check acrylic/glass lid for cracks or chips
|
Completed
|
|
Inspect vacuum chamber seal for tears or cracks
|
Completed
|
|
Confirm chamber lid micro switches report .04 Ohms
|
Completed
|
|
Verify pump or cold trap connections are sealed and leak-free
|
Completed
|
|
Inspect and clean internal fan and condenser (if applicable)
|
Completed
|
|
Inspect internal wiring for frays or breakages
|
Completed
|
|
Test motor functionality, replace motor/bearings if needed
|
Completed
|
|
Inspect internal pump seals, replace if needed
|
Completed
|
|
Inspect all internal hoses for leaks, ensure clean seals
|
Completed
|
|
Confirm unit achieves specified temperature ranges
|
Completed
|
|
Verify unit reaches specified vacuum levels
|
Completed
|
|
Conduct a 6-8 hour cycle test for operational performance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|